- 39 -
ERROR
When the Mire Image is not clear because
the cornea is dry.
Instruct the examinee to open and close his or
her eyes several times and measure again.
Examinee has strong irregular astigmatism
or corneal disease.
Impossible to measure
+ OUT
Sum of SPH and CYL of examinee’s eye is
over
+25D
.
Impossible to measure
Radius of curvature is over
13.0mm
.
Objective glass in the measurement window
is dirty
Clean the glass
- OUT
Sum of SPH and CYL of examinee’s eye is
over
-30D
.
Impossible to measure
Radius of curvature is under 5.0mm.
Objective glass in the measurement window
is dirty
Clean the glass
C OUT
Astigmatism is over 10D.
Impossible to measure
Corneal stigmatism is over 15D.
Objective glass in the measurement window
is dirty.
Clean the glass
(3) Message On Printing
Message
Cause
Method to deal with
No Print Data
There is no measurement data
Print after measurement
PRINTER PAPER EMPTY
No printing paper
Load a roll of printing paper.