PRK-8000 Operation Manual
- 28 -
(2) Ready for Observation
Ask the examinee to look at the red roof of the eye fixation target.
See procedure (3) of the Continuous Keratometry and Refractometry [K/R Mode].
(3) Observation
z
Adjust the amount of light of each light source so the image can be seen clearly.
Press one of buttons for each light source on the side of display. The light source is
adjustable from 0 to 100. If it is set to 0, the light turns off.
¾
RETRO LED: The observation light source
¾
CHART LED: Inside Fixed target light the patient sees
¾
EXTERNAL LED: A light that turns around the eyes
z
Position the light for retroillumination reflected from cornea to a part that is not
opaque by using the operation lever. If the light is directed from the edge of the pupil,
it will be easier to observe the image. Focus on the image.
z
After confirming that the image is focused, press measuring button. The image will
freeze.
z
Because the LED light in ILLUM mode is more brighter then normal measurement
mode, if you observe a long period of time it can give stress in the eye of the
examinee.
z
In the RETRO LED ON status, if there is no operation for more than 30 seconds, the
RETRO LED will be OFF automatically.
(4) Displaying of stored Image
z
Press DISP button in observation display. Images stored in memory will be
displayed.
[Menu Display]
z
When you press the Image button, selecting Image will be displayed.
Images of eye
Change Images
of eye
Serial number – Sub number
Date & Time of
the selected image
End of the
Display Mode
Clear stored Image
& Measurement
result