background image

Poseidon SE7EN User Manual

Appendix 3 Page 98

Before reaching this test (indeed, before Powering-up the electronics), it’s important to make 
sure that the over-pressure relief valve on the bottom of the right (exhale) counterlung is 
adjusted to the full clockwise position. As mentioned previously, the mouthpiece should be in 
the OC position, and the oxygen cylinder should be turned on. Also, the counterlungs should 
be no more than half inflated and the over pressure valve (OPV) should be fully closed (turned 
clock wise).

IMPORTANT: 

Make sure the over-pressure relief valve on the bottom of the right 
(exhale) counterlung is adjusted to the maximum cracking pressure 
(turned all the way in the clockwise direction). Also ensure that 
the mouthpiece is in the OC position, that the oxygen cylinder 
valve is turned on, and the counterlungs are not already inflated. 
Otherwise, Test 49 will fail. Make sure that nothing pushes against 
the OPV, that will cause the Test 49 to fail.

Figure 2-9.

  

Test 49 - Positive pressure loop test.

The sequence of events for Test 49, and the various corresponding tests that are conducted, 
are as follows. First, the system injects oxygen into the breathing loop until the counterlungs 
are both full, but not tight (detected by the depth sensor as a slight pressure increase). This is 
why it’s important that the counterlungs not already be fully inflated prior to starting Test 49 
(which can occur if the cylinders are turned on and the mouthpiece is in the OC position during 
tests 24–27). This initial inflation is done via one of the two metabolic oxygen solenoid valves, 
thereby ensuring that this solenoid valve is actually injecting gas when it is supposed to.

Once the counterlungs are fully inflated and the depth sensor detects a slight increase in 
pressure, the system pauses and monitors internal loop pressure for 20 seconds. If any of the 
four solenoid valves are leaking, the pressure inside the breathing loop will gradually rise. 
Assuming no increase in loop pressure is detected during this 20-second period, the second 
metabolic solenoid valve is used to inflate the breathing loop to a higher internal pressure. 
When this happens, the counterlungs will be tightly inflated, and the internal pressure should 
be slightly less than the cracking pressure of the over-pressure relief valve on the bottom of the 
right counterlung, when that valve is adjusted to its maximum cracking pressure. The system 
then monitors the loop pressure for the next 20 seconds to determine whether the pressure 
decreases, as by a leak in the breathing loop. 

Battery power verification (Test 48)

Immediately after passing the two gas pressure tests, the pre-dive routine tests whether there 
is sufficient battery power to begin a dive (Test 48). The amount of power required depends on 
how recently the battery was subjected to a Learn Cycle during charging (see Chapter 1). If 
the Learn Cycle occurred recently, then the system is able to predict the remaining battery life 
relatively accurately, and Test 48 will pass if the battery has at least 20% charge remaining 
(approximately 5–6 hours of typical dive time, or 4 hours of night-diving time). The amount of 
charge required to pass this test increases by 0.5% per day since the last Learn Cycle, such 
that after 160 days with no Learn Cycle, Test 48 will not pass.

This test will pass or fail immediately. If it fails, the only remedies are to re-charge the battery 
(and/or subject it to a Learn Cycle), or replace the battery with another one with greater charge 
(subject to decompression data discrepancies, as discussed previously for Test 40).

Positive pressure loop test (Test 49)

One of the most basic pre-dive tests for any rebreather is to make sure that the breathing loop 
is intact and not leaking. Water entry into the breathing loop can cause serious problems if 
mixed with the CO

2

 absorbent material in the cartridge. As discussed earlier in this Appendix, 

a manual negative pressure loop test can help detect leaks in the breathing loop. 

Another common test is the Positive Pressure Loop Test (PPLT), which is similar to the Negative 
Pressure Loop Test, except the test is performed by pressurizing the breathing loop with 
positive pressure. Like the Negative Pressure Loop test, this test could very easily be performed 
manually. However, one of the features of the Poseidon SE7EN – the placement of the depth 
sensor within the breathing loop – allows this test to be performed automatically (Figure 2-9). 
Test 49 actually performs four separate tests, only one of which is the PPLT. The other three 
are:

•  Verifies depth sensor is sensitive to small pressure changes 

•  Verifies that both metabolic oxygen solenoid valves are injecting gas 

•  Tests for leaks in all four solenoid valves.

Summary of Contents for SE7EN

Page 1: ...SE7EN connected to M28 POSEIDON SE7EN Appendix 3 version 1 2...

Page 2: ...m allowed working pressure it holds 612 liter 21 6 cubic feet of air Because this cylinder is your open circuit OC bailout gas in the event of an emergency Poseidon strongly recommends that this cylin...

Page 3: ...loop Repeat this procedure several times until you have pulled as strong a vacuum on the breathing loop as you can and then quickly switch the mouthpiece to OC position to hold the vacuum inside the...

Page 4: ...on M28 then the first screen displayed shows the SE7EN mode display Figure 2 2 for a few seconds before switching automatically to initiate the PreDiveTest mode Figure 2 3 The firmware version number...

Page 5: ...PSTs is as follows Test 1 Confirms the main data logger is functional Tests 2 9 verify the ROM RAM and EEPROM function in all four processors Test 10 verify the Bluetooth functionality Test 14 Confir...

Page 6: ...The system allows the diver up to two minutes to make this switch The Closed Circuit icon will continue to display until Test 43 when the mouthpiece needs to be switched to the open circuit OC positio...

Page 7: ...two decompression states do not exactly match exactly test 40 will fail Failure of this test is a notification to the diver that the system has detected this discrepancy between the two sets of decom...

Page 8: ...the off position when Test 44 is reached if not gas will be wasted during Tests 24 27 which verify that the four solenoid valves draw the correct amount of power when held open Each of these two test...

Page 9: ...racking pressure of the over pressure relief valve on the bottom of the right counterlung when that valve is adjusted to its maximum cracking pressure The system then monitors the loop pressure for th...

Page 10: ...mouthpiece is in the Closed Circuit position and is detected by the HUD the test will pass The system allows 2 minutes for this test to be completed before timing out As with the Open Circuit mouthpie...

Page 11: ...the RBConfig PC software Prebreathe Figure 2 13 Prebreathe Open circuit regulator function Test 54 After completing Test 53 the information field will have the text Switch mouthpiece to Open Circuit...

Page 12: ...mouthpiece tighten Attach HUD on to mouthpiece Attach LP diluent supply hose to inflator Close OPV on right counterlung Negative loop pressure test Insert smart battery and conduct power up self test...

Page 13: ...position NOW In most cases this signal will be triggered in association with an open circuit bailout situation thereby instructing the diver to switch the mouthpiece from closed circuit mode to open c...

Page 14: ...to alert other nearby divers of a potential problem Monitoring the M28 primary display Most of the information concerning the status of the dive and the various system parameters is communicated to t...

Page 15: ...etail Figure 3 2 LCD display layout Figure 3 3 Symbols DANGER If the Primary Display screen is ever blank while diving the Poseidon SE7EN immediately commence an abort to the surface in Open Circuit m...

Page 16: ...r of the screen means all systems are functioning properly and all parameters are working cor rectly In most cases the signals will flash when activated further drawing attention Abort and open circui...

Page 17: ...ive should be termi nated whenever this icon is displayed The diver should ascend towards the surface at a slow and controlled rate watching the Primary Display for the Stop Alert and additional decom...

Page 18: ...Chapter 2 the linearity of the oxygen sensor response is only validated up to a PO2 value of 1 0 bar atm i e 100 oxygen at sea level Most rebreathers assume that the sensor response remains linear at...

Page 19: ...C0 C1 C2 Mouthpiece position The difference between NC no circuit and UN unknown depends on whether the mouthpiece is reporting that neither closed circuit nor open circuit is currently established no...

Page 20: ...21 Current depth Maximum depth Ceiling Figure 3 22 Maximum depth The maximum depth achieved during the dive is displayed above the current depth in the lower left part of the Primary Display Ceiling i...

Page 21: ...26 Elapsed dive time The number of minutes that have elapsed during the dive i e the total dive time is dis played in the lower right part of the Primary Display above the RDT value This value repre...

Page 22: ...icator If the battery fails the entire life support system including alarms may cease to function Failure to abort to open circuit and terminate the dive could lead to serious injury or death Cylinder...

Page 23: ...ltering the center icon see figure 3 35 This selection will remain as long as you stay in the OC mode If you decide to go back to CC the gas will automatically be restored to your SE7EN predefined dil...

Page 24: ...t the Air Dil pressure graph is not flashing The electronics system will constantly calculate whether there is enough air supply remaining to allow a safe open circuit bailout to the surface If there...

Page 25: ...tments Besides the three large straps for attachment to the harness each counterlung has several additional straps used to adjust positioning At the top of each counterlung is a single adjust able str...

Page 26: ...specially deep inhaled breath For minor adjustments in buoyancy it is usually easier to control it with your breath rather than the purge button since it could easally give a bit too much gas To decre...

Page 27: ...following four conditions have all been met the depth is 0 the back of the Primary Display where the wet switch contacts are located has been dried the pressure in the diluent regulator and hoses has...

Page 28: ...l continue to flash and there may be a symbol in the upper right field of the display flashing to indicate the nature of the problem For example the directional arrows advise you to go up ascend if yo...

Page 29: ...the physical product needs to be connected and then the Power out option needs to be enabled To enable the power out function goto the SYSTEM app and press both buttons Select Connect and then Power...

Page 30: ...1 6 l CO2 per minute at STPD Standard Temperature Pressure Dry The SE7EN can support decompression diving To be able to perform decompression dives you need two things 1 the appropriate training and 2...

Page 31: ...of the counterlungs to the gas processor Figure 1 16 Three velcro straps on the back of each counterlung Figure 1 16 attach the counterlungs to the shoulder straps of the backpack The SE7EN is provid...

Page 32: ...s These valves can be replaced with blank covers if manual gas addition capabilities are not required e g for recreational diving The port for the manual Oxygen addition valve right counterlung is int...

Page 33: ...tor hose to your onboard diluent gas supply or any other offboard source Route the hose over your left sholder and connet to the diluent inflator valve though the quick connection The valves described...

Page 34: ...has been reached The generated alarm response will be similar to the alarm shown when the diver exceeds the maximum working depth of the rig This value also affects the way the controlling resource al...

Page 35: ...the diluent composition to air If the Recreational 40m battery or Deco 40m battery was inserted by mistake the diver should NOT confirm these tests Instead the test should be allowed to time out two m...

Page 36: ...ill not issue an alarm HUD Buddy lights Audio Alarm when decompression is entered There is no audio alarm when decompression is entered on a non decompres sion configuration When the mouthpiece is fou...

Reviews: