background image

Poseidon SE7EN User Manual

Appendix 3 Page 115

Tips on breathing

Breathing underwater on a closed-circuit rebreather, such as the Poseidon SE7EN, is 
somewhat different from breathing on land, or breathing with conventional scuba gear. As the 
diver exhales, the counterlungs both expand. As the diver inhales. the counterlungs contract. 
The direction of gas flow through the breathing loop is governed by the two check-valves in 
the bottom portion of the mouthpiece. The incorporation of two separate, over-the-shoulder 
counterlungs on the SE7EN helps to minimize the effort required to breathe underwater, but 
there are a few tips that make breathing easier.

The most important thing is to maintain an optimum volume of gas in the breathing loop. If 
there is too much back-pressure when exhaling (often felt in the cheeks), or if the overpressure 
relief valve on the exhale (left) counterlung “burbs” gas at the end of an exhaled breath, then 
the loop has too much gas, and some should be vented (e.g., by exhaling through the nose). 
If the counterlungs “bottom out” and/or the Automatic Diluent Valve (ADV) in the mouthpiece 
is triggered on a full inhalation, then there is not enough gas in the breathing loop. This condition 
should be corrected automatically by the ADV.

Tips on buoyancy control

Controlling buoyancy while diving with a rebreather is considerably different from buoyancy 
control with conventional open-circuit scuba. To begin with, whereas a scuba diver needs to 
manage buoyancy characteristics of two separate factors: the Buoyancy Control Device 
(BCD), and the exposure suit (i.e., a wetsuit or a dry suit). A rebreather diver must manage 
both of these, as well as the breathing loop of the rebreather. A complete discussion of 
buoyancy control with closed-circuit rebreathers is beyond the scope of this Manual. However, 
the following tips might be useful.

Although most divers probably do not realize it, fine trim for diving with conventional scuba 
gear is achieved through breathing. On each inhalation, the diver’s lungs expand and buoyancy 
is increased. The opposite occurs on exhalation. However, this does not occur with a rebreather 
(the Poseidon SE7EN included), because the buoyancy increase caused by expanding the 
lungs on an inhaled breath is offset by the decreasing volume of the counterlungs (and vice 
versa). This may at first be disconcerting for an experienced scuba diver trying a rebreather for 
the first time, because an inhalation done subconsciously to slightly increase buoyancy has no 
effect. However, with practice, it becomes advantageous to be able to hover in the water with 
perfect buoyancy, while breathing continuously.

The quickest and easiest way to fine-tune buoyancy with a rebreather is via addition and 
removal of gas to or from the breathing loop. To increase buoyancy slightly, a small amount of 
gas can be added to the breathing loop via the ADV (either by manually engaging the purge 
button, or by making an especially deep inhaled breath). For minor adjustments in buoyancy, 

it is usually easier to control it with your breath rather than the purge-button since it could 
easally give a bit too much gas. To decrease buoyancy slightly, one need only exhale through 
the nose to vent gas out of the breathing loop (except when certain kinds of full-face masks 
are used).

New rebreather divers often have the most difficulty in very shallow water, where a slight 
change in depth yields a proportionally large change in displacement (and, hence, buoyancy). 
This is especially true when the diver begins to ascend, which causes the counterlungs to 
expand, leading to increased buoyancy, leading to further ascents, and expanding loop 
volume. This can lead to a “run-away” ascent that can be difficult to control. For this reason, 
it’s useful practice for rebreather divers to be in the habit of venting gas through the nose 
whenever ascending; particularly from very shallow depths.

Venting water from the loop

Even if a diver is very careful to prevent water from entering the breathing loop, there will always 
be some water collecting due to condensation. Most of this will form on the “exhalation” side 
of the breathing loop, between the mouthpiece and the CO

2

 absorbent cartridge, and will 

generally collect in the exhalation (right-hand) counter lung. Sometimes, water will collect in the 
exhalation hose, immediately downstream of the mouthpiece. If this water is sufficient to cause 
gurgling noises with each breath, it can be poured into the exhalation counter lung by looking 
upward and holding the hose in such a way so as to dump the water towards the right-hand 
shoulder port. In most cases, the water that collects inside the exhalation counter lung will not 
disrupt the function of the Poseidon SE7EN in any way, so it can be safely ignored. However, 
sufficient quantities of water could be returned to the breathing loop if the diver becomes 
inverted, so it may be desirable to vent this water from the breathing loop altogether.

To do this, the diver should first become negatively buoyant, or attach to a secure object on 
the bottom. The breathing loop volume should be increased to at least 75% of maximum 
capacity by manually adding diluent via the ADV. The loop vent valve at the bottom of the 
exhalation counter lung should be rotated counterclockwise maximally to minimize the cracking 
pressure. While in an upright orientation, the diver should then compress both counter lungs 
by squeezing them against the chest with the elbows and upper arms, while simultaneously 
exhaling through the mouth and depressing the loop vent valve to open it. If done correctly, 
water will be expelled from the loop vent valve first, followed by a stream of gas bubbles. After 
the water has been flushed, the loop vent valve can be tightened by rotating clockwise, and 
the breathing loop volume and PO

2

 can be restored to normal.

A small amount of condensation may also collect in the inhalation portion of the breathing loop, 
between the CO

2

 absorbent cartridge and the mouthpiece. Normally, this will only be a small 

volume of water, and most will be absorbed by water trap.

Summary of Contents for SE7EN

Page 1: ...SE7EN connected to M28 POSEIDON SE7EN Appendix 3 version 1 2...

Page 2: ...m allowed working pressure it holds 612 liter 21 6 cubic feet of air Because this cylinder is your open circuit OC bailout gas in the event of an emergency Poseidon strongly recommends that this cylin...

Page 3: ...loop Repeat this procedure several times until you have pulled as strong a vacuum on the breathing loop as you can and then quickly switch the mouthpiece to OC position to hold the vacuum inside the...

Page 4: ...on M28 then the first screen displayed shows the SE7EN mode display Figure 2 2 for a few seconds before switching automatically to initiate the PreDiveTest mode Figure 2 3 The firmware version number...

Page 5: ...PSTs is as follows Test 1 Confirms the main data logger is functional Tests 2 9 verify the ROM RAM and EEPROM function in all four processors Test 10 verify the Bluetooth functionality Test 14 Confir...

Page 6: ...The system allows the diver up to two minutes to make this switch The Closed Circuit icon will continue to display until Test 43 when the mouthpiece needs to be switched to the open circuit OC positio...

Page 7: ...two decompression states do not exactly match exactly test 40 will fail Failure of this test is a notification to the diver that the system has detected this discrepancy between the two sets of decom...

Page 8: ...the off position when Test 44 is reached if not gas will be wasted during Tests 24 27 which verify that the four solenoid valves draw the correct amount of power when held open Each of these two test...

Page 9: ...racking pressure of the over pressure relief valve on the bottom of the right counterlung when that valve is adjusted to its maximum cracking pressure The system then monitors the loop pressure for th...

Page 10: ...mouthpiece is in the Closed Circuit position and is detected by the HUD the test will pass The system allows 2 minutes for this test to be completed before timing out As with the Open Circuit mouthpie...

Page 11: ...the RBConfig PC software Prebreathe Figure 2 13 Prebreathe Open circuit regulator function Test 54 After completing Test 53 the information field will have the text Switch mouthpiece to Open Circuit...

Page 12: ...mouthpiece tighten Attach HUD on to mouthpiece Attach LP diluent supply hose to inflator Close OPV on right counterlung Negative loop pressure test Insert smart battery and conduct power up self test...

Page 13: ...position NOW In most cases this signal will be triggered in association with an open circuit bailout situation thereby instructing the diver to switch the mouthpiece from closed circuit mode to open c...

Page 14: ...to alert other nearby divers of a potential problem Monitoring the M28 primary display Most of the information concerning the status of the dive and the various system parameters is communicated to t...

Page 15: ...etail Figure 3 2 LCD display layout Figure 3 3 Symbols DANGER If the Primary Display screen is ever blank while diving the Poseidon SE7EN immediately commence an abort to the surface in Open Circuit m...

Page 16: ...r of the screen means all systems are functioning properly and all parameters are working cor rectly In most cases the signals will flash when activated further drawing attention Abort and open circui...

Page 17: ...ive should be termi nated whenever this icon is displayed The diver should ascend towards the surface at a slow and controlled rate watching the Primary Display for the Stop Alert and additional decom...

Page 18: ...Chapter 2 the linearity of the oxygen sensor response is only validated up to a PO2 value of 1 0 bar atm i e 100 oxygen at sea level Most rebreathers assume that the sensor response remains linear at...

Page 19: ...C0 C1 C2 Mouthpiece position The difference between NC no circuit and UN unknown depends on whether the mouthpiece is reporting that neither closed circuit nor open circuit is currently established no...

Page 20: ...21 Current depth Maximum depth Ceiling Figure 3 22 Maximum depth The maximum depth achieved during the dive is displayed above the current depth in the lower left part of the Primary Display Ceiling i...

Page 21: ...26 Elapsed dive time The number of minutes that have elapsed during the dive i e the total dive time is dis played in the lower right part of the Primary Display above the RDT value This value repre...

Page 22: ...icator If the battery fails the entire life support system including alarms may cease to function Failure to abort to open circuit and terminate the dive could lead to serious injury or death Cylinder...

Page 23: ...ltering the center icon see figure 3 35 This selection will remain as long as you stay in the OC mode If you decide to go back to CC the gas will automatically be restored to your SE7EN predefined dil...

Page 24: ...t the Air Dil pressure graph is not flashing The electronics system will constantly calculate whether there is enough air supply remaining to allow a safe open circuit bailout to the surface If there...

Page 25: ...tments Besides the three large straps for attachment to the harness each counterlung has several additional straps used to adjust positioning At the top of each counterlung is a single adjust able str...

Page 26: ...specially deep inhaled breath For minor adjustments in buoyancy it is usually easier to control it with your breath rather than the purge button since it could easally give a bit too much gas To decre...

Page 27: ...following four conditions have all been met the depth is 0 the back of the Primary Display where the wet switch contacts are located has been dried the pressure in the diluent regulator and hoses has...

Page 28: ...l continue to flash and there may be a symbol in the upper right field of the display flashing to indicate the nature of the problem For example the directional arrows advise you to go up ascend if yo...

Page 29: ...the physical product needs to be connected and then the Power out option needs to be enabled To enable the power out function goto the SYSTEM app and press both buttons Select Connect and then Power...

Page 30: ...1 6 l CO2 per minute at STPD Standard Temperature Pressure Dry The SE7EN can support decompression diving To be able to perform decompression dives you need two things 1 the appropriate training and 2...

Page 31: ...of the counterlungs to the gas processor Figure 1 16 Three velcro straps on the back of each counterlung Figure 1 16 attach the counterlungs to the shoulder straps of the backpack The SE7EN is provid...

Page 32: ...s These valves can be replaced with blank covers if manual gas addition capabilities are not required e g for recreational diving The port for the manual Oxygen addition valve right counterlung is int...

Page 33: ...tor hose to your onboard diluent gas supply or any other offboard source Route the hose over your left sholder and connet to the diluent inflator valve though the quick connection The valves described...

Page 34: ...has been reached The generated alarm response will be similar to the alarm shown when the diver exceeds the maximum working depth of the rig This value also affects the way the controlling resource al...

Page 35: ...the diluent composition to air If the Recreational 40m battery or Deco 40m battery was inserted by mistake the diver should NOT confirm these tests Instead the test should be allowed to time out two m...

Page 36: ...ill not issue an alarm HUD Buddy lights Audio Alarm when decompression is entered There is no audio alarm when decompression is entered on a non decompres sion configuration When the mouthpiece is fou...

Reviews: