E-4
E Interface Operation RS-232 and IEEE-488 / GPIB
Displacement Decoder
Examples:
Command
Description
Answer
Answer with
ECHOON
AMPL?
Queries the set measurement
range
Range number
(ASCII character)
1 ... 7
(refer to table below)
AMPL1 ...
AMPL7
AMPL1 ...
AMPL7
Sets the corresponding
measurement range (refer to
table below)
-
AMPL1 ...
AMPL7
AMPL
Loads the initialization settings
for the measurement range
(refer to command RENDCL)
-
AMPL
RES
Triggers the reset function of the
displacement decoder
-
RES
Measurement range
Range number
µ
m/ V
OVD-10
OVD-20
0.5
-
0
2
-
1
8
-
2
20
3
3
80
4
4
320
5
5
1,280
6
6
5,120
7
7
AMPL2\n
If an OVD-20 is installed, the measurement range is set to
8
µ
m / V.
AMPL7\n
The measurement range is set to 5,120
µ
m / V.
AMPL?\n
As an answer, a string with the number of the displacement
measurement range set is returned.
If the measurement range 320
µ
m / V is selected, the controller
returns the string "5\n".