7-1
7 Technical Specifications
7 Technical Specifications
7.1 Controller OFV-3001
7.1.1 General Data
Standards Applied
7.1.2 Interfaces
Mains voltage:
100 / 115 / 230 VAC
±
10%, 50 / 60 Hz,
adjustable at the back panel
Power consumption:
max. 150 VA
Fuses:
1.0 A / slow-blow for 230 V
2.0 A / slow-blow for 100 / 115 V
Protection class:
I (protective grounding)
Operating temperature:
+
5°C...
+
40°C (41°F... 104°F)
Storage temperature:
−
15°C ...
+
65°C (5°F... 149°F)
Relative humidity:
max. 80%, non-condensing
Dimensions:
450 mm
×
355 mm
×
135 mm
Weight:
10.8 kg
Calibration recommended:
every 2 years
Electrical safety:
EN 60950 (IEC 950), EN 61010 (IEC1010)
EMC:
Emission: EN50081-1 (FCC Class B)
Immunity: EN50082-1, EN 50082-2 (IEC801-1...-5)
Laser safety:
EN 60825-1 (CFR 1040.10, CFR1040.11)
RS-232:
8 data bits, no parity, baud rate 4,800 or 9,600
9-pin female Sub-D cable to the workstation,
1:1 wired
IEEE-488 / GPIB:
according to IEEE-488.1
REMOTE FOCUS:
special interface for the hand terminal OFV-310
EXT. DEC.:
special interface for the PC-based displacement
decoder VibSoft FC
SIGNAL:
0 V... 3 V DC, proportional to the logarithm of the
optical signal level, load resistance
≥
10 k
Ω