3-15
3 First Steps
Test
10. Put a piece of reflective film (enclosed in this manual) at approximately
12 cm from the front lens of the mini sensor in the beam path.
11. Focus the laser beam on the reflective film by turning the mini sensor.
Providing the sensor head and the input section of the controller are
working correctly, the signal level display will fully light up.
12. If you have not been able to observe the effect described under
., check
the signal level again after 20 minutes. After this warm-up phase the laser
has reached its working temperature (refer also to
If the functional test has been successful you can now make measurements
as described in
If your vibrometer does not perform as described above, read through the
information on fault diagnosis provided in
and, if necessary, contact
your local Polytec representative.