12 DiTechnical details
Dimensions Casing
display
780
[mm]
Width
380
[mm]
Depth (casing)
830
[mm]
Do not operate backside
directly to the wall
Weight
73
[kg]
Power consumption
Initial start (first 15 min. max)
2600W
During first
starting
minutes
Cooling down
Working load
1500W max
During normal operation and
no heating
Heater
1400W max
Full operation
2900W max
Heating up mode
Working conditions
Temperature
15..26
[°C]
Humidity
10..90
[%]
relative humidity
Bath
Opening
175x130
[mm]
Depth
260
[mm]
Contents
11
[ltrs]
Heat removal capacity
@ -84°C / -119 F
250
[Watt]
@ -82°C / -116 F
300
[Watt]
@ -73°C / -99 F
350
[Watt]
@ -64°C / -83 F
400
[Watt]
@ -58°C / -72 F
450
[Watt]
Cooling down curve
TCC-B Cooldown Speed
+20
0
Tem
per
a
ture
[
°C]
-80
-88
0:00
0:30
1:00
1:10
1:30
Time [hrs:min]