7.1 Fluid level
Do not operate the bath with low fluid level.
When the fluid level is too low, bath fluid will vaporize leading
to toxic and flamable fumes.
Flamable fumes can lead to fire
Flammable fumes can be ignited by the not submersed part
of the heating element.
The level indicator will start to blink (blue light) when the
fluid level is too low.
When the bath has been installed it must be filled with an
appropriate liquid. Depending on the operating temperature
the liquid level in the bath should be observed.
The liquid level should be maintained between 1 and 3 cm
below the lid during normal operation.
The heating element will be damaged when not fully
submerged in the bath fluid. A lower level than 5 cm below the
lid may damage the heaters. A high bath level can cause
overflow and will might also damage the bath insulation.
Low fluid level:
-
Can cause fire when heater gets partially submersed
-