7.4 Start
The bath must be filled with a liquid suitable for the minimum operating temperature.
It is very important to select a liquid with a viscosity of
less than 20 cSt at the operating temperature and a flash
point which is well above the operating temperature.
The best bath liquid to be used is methanol, enabling
circulation at very low temperatures.
Methanol is extremely flammable and can cause fire hazard.
Please take all necessary precautions to reduce fire hazard.
When using methanol all vapors must be removed by using
appropriate air ventilation.
Methanol is toxic and can cause health risks. Use appropriate
ventilation and other precautions to prevent inhaling toxic
vapors. If ventilation is insufficient the risk of explosion
hazards can occur!
The supplier of the bath liquid (methanol) will be able to hand
over all chemical details and safety precautions related to the
use of methanol. These precautions must be followed when
operating the bath.
VISCOSITY
< 20 cSt
FLASHPOINT
> OPERATING
TEMPERATURE
METHANOL