CHAPTER 1: QUICK START 

1.1 The Quickest Start Possible 

Your miniDOT Clear Logger has arrived completely ready to go. It is set to measure and record time, 
battery voltage, temperature, oxygen concentration, and measurement quality once every 10 
minutes and write one file of measurements daily. Open the miniDOT Clear Logger and move the 
Logger Control Switch to the “Record” position. In this condition, the miniDOT Clear Logger will 
record measurements for a year before the internal batteries are expended. You must re-close the 
miniDOT Clear Logger prior to deploying it.  

At the end of the deployment period, open the miniDOT Clear Logger and connect it to a HOST 
computer via the USB connection. 

The miniDOT Clear Logger will appear as a ‘thumb drive’. Your 

temperature and oxygen concentration measurements, together with a time stamp indicating the 
time the measurements were made, are recorded in text files in the folder having the serial number 
of your miniDOT Clear Logger. These files can be copied onto any Windows or Mac HOST 
computer. 

This Manual and other software programs are also recorded on the miniDOT Clear Logger.  

 

MINI

DOT

 

C

ONTROL PROGRAM

:

 

Allows you to see the state of the miniDOT Clear Logger as 

well as set the recording interval.

 

 

 

MINI

DOT

 

P

LOT PROGRAM

:

 

Allows you to see the plots of the recorded measurements. 

 

MINI

DOT

 

C

ONCATENATE PROGRAM

:

 

Gathers all the daily files into one CAT.txt file.

 

Your miniDOT Clear Logger will return to recording measurements after you disconnect the USB 
connection. If you wish to stop recording, then move the Logger Control Switch to t

he “Halt” position. 

You may move the Logger Control Switch at any time.  

Follow these steps to start the deployment, logging DO & T once each 10 minutes: 

1.  Open the miniDOT Clear Logger by unscrewing the clear pressure housing from the black 

end cap. It opens like a flashlight. Remove the clear pressure housing completely. Inside you 
will see the circuit pictured below: 

Summary of Contents for miniDOT Clear

Page 1: ...SUPPORT INFO PME COM 760 727 0300 PG 1 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED 1483 POINSETTIA AVE STE 101 VISTA CA 92081 USA miniD...

Page 2: ...ART 6 1 1 The Quickest Start Possible 6 1 2 A Few Details 8 1 3 Overview General Maintenance 9 CHAPTER 2 SOFTWARE 15 2 1 Overview and Software Installation 15 2 2 miniDOT Control 16 2 3 miniDOT Plot 1...

Page 3: ...excludes CT C sensors 1 year C Sense Logger Logger only 1 year For valid warranty claims made and covered defects existing during the applicable warranty period PME will at PME s option repair replac...

Page 4: ...sible or liable to purchaser or otherwise for any indirect incidental special exemplary or consequential damages including but not limited to lost profits loss of data loss of use business interruptio...

Page 5: ...into contact with the enclosed batteries then the batteries may generate gas causing the internal pressure to increase This gas will likely exit via the same location where the water entered but not n...

Page 6: ...ation measurements together with a time stamp indicating the time the measurements were made are recorded in text files in the folder having the serial number of your miniDOT Clear Logger These files...

Page 7: ...loyment 1 Recover the miniDOT Clear Logger 2 Clean and dry all accessible surfaces except the sensing foil 3 Open the miniDOT Clear Logger by unscrewing the clear pressure housing from the black end c...

Page 8: ...r Logger measures and records time battery voltage temperature dissolved oxygen concentration and measurement quality at equal time intervals The default time interval is 10 minutes However it is also...

Page 9: ...allocate new file space A few hundred files on the SD card is not a problem but as the number of files grows large into the thousands then the miniDOT Clear Logger may suffer decreased battery life o...

Page 10: ...de batteries Sample Interval Minutes Main AA Battery Life Months Number of Samples 1 12 500K 10 12 52 000 60 12 8 000 Keep a general record of the miniDOT Clear Logger s number of samples It is not po...

Page 11: ...ttery may leak SEE SECTION 3 4 FOR CAUTION ON BATTERY PLACEMENT Error on the side of caution when planning your deployment The recommended battery is the Energizer L91 lithium battery For more informa...

Page 12: ...deposited on the inner surface of the o ring When the clear pressure housing is screwed back on to the black end cap then these drops can become trapped inside the miniDOT Clear Logger Be sure to care...

Page 13: ...ced to sea level You must determine the absolute barometric pressure at your elevation A more comprehensive experiment is to additionally place ice in the bucket and mix until the water temperature is...

Page 14: ...WHEN NOT IN USE Remove the batteries Keep the black end covered with the cap supplied by PME If the cap is lost then cover the black end cap with aluminum foil There may be a calibration effect of am...

Page 15: ...ctory of the miniDOT Clear Logger PME suggests you leave these programs where they are on the miniDOT Clear Logger but you may copy them to any folder on your HOST computer s hard drive miniDOT Contro...

Page 16: ...will be displayed If more than a week has passed since time was last set then the miniDOT Clear Logger s clock will be set and the check mark icon will appear If the HOST computer is not connected to...

Page 17: ...er water salinity Select the folder that contains the files recorded by the miniDOT Clear Logger If the miniDOT Plot program is run directly from the miniDOT Clear Logger then the program will suggest...

Page 18: ...one session of the program In this case the software produces multiple plots Unfortunately the plots are presented exactly on top of one other and so when a new plot appears it is not obvious that th...

Page 19: ...If Surface Elevation is entered then no compensation for weather induced barometric pressure variation is made Enter elevation or barometric pressure Enter water salinity Select the folder that conta...

Page 20: ...E COM 760 727 0300 PG 20 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED The CAT txt file will resemble the following End the miniDOT Conca...

Page 21: ...ening a flashlight Turn the clear pressure housing counterclockwise relative to the black end cap Close the miniDOT Clear Logger by reversing this procedure after being sure that the o ring is free fr...

Page 22: ...anual The Logger Control Switch controls the miniDOT Clear Logger s mode Record When the switch is in this position the miniDOT Clear Logger is recording measurements Halt When the switch is in this p...

Page 23: ...logger is set to the default 10 minute sample interval the screen will display a reading 10 minutes after the logger is set to Record mode At this time the logger will display the latest temperature...

Page 24: ...o indicate that the software is beginning operation within a second or two after you complete the battery installation At this time the logger will enter the mode selected by the Logger Control Switch...

Page 25: ...AR LOGGER 1 Use a Phillips screwdriver to remove 3 of the 6 screws every other one Do NOT remove all screws At least 3 must always remain screwed in 2 Position the Copper Plate faced down so that the...

Page 26: ...d against miniDOT Clear s sensing foil they will bias the miniDOT Clear s measurement miniDOT Clear s sensing end is heavy compared to the rest of the instrument miniDOT Clear will therefore tend to h...

Reviews: