C L O S I N G A N D O P E N I N G
Close and open the miniDOT Clear Logger like you would a flashlight; open by unscrewing the clear
pressure housing from the black end cap. Close by screwing the clear pressuring housing on to the
black end cap. When closing, do not tighten the clear pressure housing. Just screw it on until it
contacts the black end cap. See Chapter 3 for more instructions.
Caution: DO NOT remove the stainless-steel screws in the black end cap. There are no user serviceable parts here.
If the screws are removed, then you will damage the miniDOT Clear Logger and it will have to be returned for repair.
S T O R A G E W H E N N O T I N U S E
Remove the batteries. Keep the black end covered with the cap supplied by PME. If the cap is lost,
then cover the black end cap with aluminum foil. There may be a calibration effect of ambient lighting
so attempt to keep ambient light from reaching the sensing foil as much as possible.
J A V A
miniDOT Clear programs depend on Java and require Java 1.7 or higher
.
Update Java at
E N V I R O N M E N T A L U S E A N D S T O R A G E C O N D I T I O N S
miniDOT Clear is useful over the range of 0 to 150% saturation of dissolved oxygen, over the range
of 0 to 35 deg C temperature and may be continuously immersed in fresh or salt water to a
maximum depth of 100 meters. miniDOT Clear may be stored in environments ranging from 0 to
100% humidity and temperatures ranging from -20 deg C to +40 deg C.
E L E C T R I C A L P O W E R S P E C I F I C A T I O N S
miniDOT Clear is battery powered and requires 2 AA size expendable or rechargeable batteries.
Voltage requirement is 3.6 VDC. Maximum current demand is 30 mA.