![Plasmapp Sterlink FPS-15s Plus User Manual Download Page 64](http://html1.mh-extra.com/html/plasmapp/sterlink-fps-15s-plus/sterlink-fps-15s-plus_user-manual_1565811064.webp)
Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
64
C2. Technical specification
Description
Unit
Dimensions
433 (W) × 614 (D) × 437 (H)
mm
Effective
chamber volume
14
liter
Cycle time
POUCH mode
SR
TM
/ SC
TM
Process
3
min
Sterilization Process
4
Overall Cycle
7
POUCH Plus mode
SR
TM
/ SC
TM
Process
6
Sterilization Process
8
Overall Cycle
14
CHAMBER mode
SR
TM
/ SC
TM
Process
21
Sterilization Process
15
Overall Cycle
36
Sterilization
Sterilant: Hydrogen peroxide (concentration: 58%)
Injection / penetration, diffusion and purification by plasma
Sterilizing
pressure
Final vacuum pressure
< 3.0
torr
Vacuum leakage
< 0.1
torr/sec
Pressure range
0.0 – 760
torr
Cycle
temperature
< 56
o
C
Display
7'' TFT LCD touch screen
Power input
100-120 / 220-240 VAC, 50/60 Hz
Max. power
consumption
1.0
kVA
Electrical
shock class
Class 1
(Protection standard and class for electrical shock)
Life cycle
15 years