9.3. User errors
The STERLINK
®
FPS-15s Plus sterilization system is designed to provide alarm messages with
the intuitive figures representing the predictable source for the user errors. When the user
confirms, the error should be checked and understood by the user. If the problems persist, please
contact Plasmapp or local Plasmapp customer service representative.
Figure 9.3 GUIs for user errors
Number
Name
Description
3
Confirm
The user should confirm the error when
the source of the problem is
checked and understood.
4
DOOR NOT
CLOSED
Probable cause: The door is not closed.
Suggestion: Check door closing and try again.
5
INCOMPLETE
SEALING
Probable cause: The STERPACK
®
and
STERPACK
®
Plus
sealing is
not properly performed.
Suggestion: Re-seal tightly and try again.
6
DESCRIPTION
Describes the causes and solutions of the problems.
“Incomplete sealing” error may occur when the pouch is damaged by sharp medical device in
POUCH mode. It is recommended to use the exclusive tray.