![Plasmapp Sterlink FPS-15s Plus User Manual Download Page 15](http://html1.mh-extra.com/html/plasmapp/sterlink-fps-15s-plus/sterlink-fps-15s-plus_user-manual_1565811015.webp)
Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
15
Sterilization process - Phase 1
∎
Injection: The hydrogen peroxide is transferred from the cassette into the vaporizer and
chamber of pouch. Specially in the Pouch mode, the injection is done directly in the pouch to
maximize the efficiency of sterilization, quick sterilization process is possible within 10
minutes.
∎
Diffusion: Pressure in the chamber or pouch is increased in order to apply hydrogen peroxide
to the surfaces of the devices and into the lumens of the load.
∎
Plasma purification / Vacuuming: The pressure within the chamber or pouch is reduced to
remove air and residual sterilant and water. The plasma power is applied to the plasma source
to purify residual hydrogen peroxide vapor. Specially in the Pouch mode, it is possible to
complete the quick sterilization process by directly removing the sterilizing agent from the
pouch.
Sterilization process - Phase 2
The steps in the
Phase 1
are repeated.