FPS-15s Plus User Manual
35
Number
Name
Description
27
Caution for door
handling
It highlights that door should not be handled by the user during the
stopping process.
28
Caution for
sterility
The status image will be blinking during the sterilization process.
29
User confirm
When the user confirm is selected, the screen moves to ready state.
Even the stopping process includes purification process, sterilant can be left on the loads. Wear
protective gloves when handling the load and used cassettes. As depicted in the final GUI, the
sterility assurance level may not be obtained, and it requires to restart a new sterilization cycle
with rewrapping using new packaging material.
WARNING! POSSIBLE NON-STERILE DEVICE.
Cycle cancellation may result in a non-sterile device, and it requires to restart a new
sterilization cycle with rewrapping using new packaging material.
CAUTION! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear latex, PVC (vinyl), or nitrile gloves whenever handling a load after a cycle
cancellation or error occurrence. Hydrogen peroxide liquid may be present on the load
or in the chamber.