![Plasmapp Sterlink FPS-15s Plus User Manual Download Page 45](http://html1.mh-extra.com/html/plasmapp/sterlink-fps-15s-plus/sterlink-fps-15s-plus_user-manual_1565811045.webp)
Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
45
7.2. Data
The STERLINK
®
FPS-15s Plus sterilizer can create, store, display, and print the cycle data.
When connected to internet, the sterilizer can upload the cycle history files to the ITS
TM
server.
Cycle history data contains very detailed sterilization cycle information of each SR
TM
,
sterilization and SC
TM
process. The data includes identifying information about the sterilizer,
time, date, cycle duration information, process data from the sensors of the sterilizer. Information
is extracted from the cycle history data to produce the printed report.
The history data is stored in the memory of the STERLINK
®
FPS-15s Plus sterilizer. The
memory holds data form the last 50 cycles. After 50 cycles are completed, the oldest cycle history
record is overwritten with new data from the 51st cycle. If the sterilizer is configured with
internet connection to ITS
TM
server, the cycle history data can be periodically uploaded to the
server.