![Planer Kryo360-1.7 Operator'S Manual Download Page 40](http://html1.mh-extra.com/html/planer/kryo360-1-7/kryo360-1-7_operators-manual_1562044040.webp)
36
Liquidus Tracker
8. A
Liquidus Tracker
button will then appear.
9. Click the
Liquidus Tracker
button to display the
Special values for PID Extended
window.
10.In the
Coefficients
section enter the appropriate parameters for the liquidus tracker. The terms
are explained in the section,
Liquidus Tracking parameters
.
Tip
If you want to retain the information for future use, click on the
To file
button to store the
parameters on your computer. They can then be retrieved by using the
From file
button.
11.When complete, click the
Send
button to transmit the values to the controller. Ensure that the
window shows a
indicating successful transmission.
Tip
It is possible to check the coefficients in the controller by clicking the
Read
button. However,
some values are derived from the controller's stored parameters and as such the returned
coefficients will not be exactly the same as originally entered. In particular, dual solution mode,
the Vr parameter will always be returned as 1 and the F factor will be modified by the same ratio as
the new Vr.
3.2.2.1
Liquidus Tracking parameters
The system is able to operate in two different modes;
Dual solution mode
and
Single solution
mode
.
In dual solution mode, two pumps are used to provide high and low concentration solutions that are
mixed before entering the sample container thereby providing the sample with cryoprotectant at the
required concentration.
36
38
40
Summary of Contents for Kryo360-1.7
Page 1: ...Liquidus Tracker Operator s Guide...
Page 5: ...Introduction Section I...
Page 20: ......
Page 21: ...Installing the system Section II...
Page 29: ...25 Installing the system...
Page 33: ...29 Installing the system 10 Connect the power supply cables to the pumps and chamber...
Page 35: ...Operating the system Section III...
Page 65: ...Routine maintenance and troubleshooting Section IV...
Page 68: ......
Page 69: ...Additional information Section V...
Page 76: ......
Page 80: ...MA101894 Issue 2 Liquidus Tracker Operator s Guide...