Function of the on-site display
The register display consists of two rows
of seven-segment digits, status, unit of
measures, flow rate, and battery
indicators.
Operating function settings and
programming are provided using the
TOTAL
and
RESET
buttons.
Normal operation
To enter normal operation mode - when the screen is blank after exiting programming
mode, or upon initial use, press either the
TOTAL
or
RESET
button once.
Status
The status indicators are RESET and TOTAL
Totalizers
The top row of indicators is the batch totalizer. This totalizer displays the cumulative
volume of flow through the meter with six digits. The batch totalizer totalizes in selected
units of measure.
To reset the batch totalizer, after 2 seconds of no flow, press and release the
RESET
button.
The bottom row of indicators display the resettable totalizer with five digits or the five least
significant digits of the non-resettable totalizer. RESET and TOTAL is indicated when the
resettable total is displayed in the five-digit lower row. Only TOTAL is indicated when the
non-resettable total is displayed.
To toggle between the non-resettable totalizer and the resettable totalizer, press and
release the TOTAL button.
To reset the resettable totalizer, press and hold the TOTAL button and then press and
release the RESET button.
To display 11-digit non-resettable totalizer, while the non-resettable total is displayed,
press and hold the TOTAL button for seconds. The top row displays the 6 most significant
digits; the bottom row displays five least-significant digits.
NOTE:
The non-resettable totalizer normally displays 5 least-significant digits.
DOZ03 / DOZ05 / DOZ07 Instruction manual 02/2021
page 17