Page 39
PIE 525Plus Specifications
Read mA
Ranges and
Resolution
0.000 to 24.000 mA or -25.00 to 125.00% of
4-20 mA
Accuracy
≤ ± (0.02 % of R 0.003 mA)
Voltage burden
≤ 2V at 24 mA
Overload/Current
limit protection
25 mA nominal
Source mA / Power & Measure Two Wire Transmitters &
PWRM LEAK
Ranges and
Resolution
0.000 to 24.000 mA or -25.00 to 125.00% of
4-20 mA
Accuracy
≤ ± (0.02 % of R 0.003 mA)
Loop compliance
voltage
≥ 24 DCV @ 20.00mA
Loop drive
capability
1200 Ω at 20 mA
950 Ω with Hart Resistor or leak detection running
mA 2-Wire Transmitter Simulation
Accuracy
Same as Source/Power & Measure
Voltage burden
≤ 2V at 20 mA
Overload/
Current limit
protection
24 mA nominal
Loop voltage
limits
2 to 60 VDC (fuse-less protected from
reverse polarity connections)
www.
GlobalTestSupply
.com
Find Quality Products Online at: