Page 17
Voltage Output Signal
Controller
Transmitter
Power Supply
mV
2 1
52.501
+
-
MAX
OFF
OVERLOAD
Double Click
Menu
HI
SET
LO
SOURCE
MAX
READ
MIN
READ
Push & Hold
to
Store/Step
Push & Turn
for
Fast Dialing
Diagnostic Thermocouple
RTD & Milliamp Calibrator
2, 3, 4 Wire Detection
Loop Leak Detect ∙ Transmitter Supply
Read mV
Choose this function to measure from -13.000 to 80.000 mV in
LOW resolution and -13.0000 to 80.0000 mV in HIGH resolution.
Move the power switch
w
to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select V for the FUNCTION.
Connect the red input lead (+) of the PIE 525Plus to the more
positive point and the black input lead (-) to the more negative
point.
Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” flashed on the display and the red
OVERLOAD LED lit.
The PIE 525Plus measures the input signal and constantly
updates the display with the current reading. Move the
EZ-CHECK switch
q
to MAX to see the highest reading and to
MIN to see the lowest reading. Press and hold the knob
e
to
clear the MAX and MIN readings.