Page 18
Source Thermocouple
Choose this function to provide a simulated thermocouple signal
into controllers, temperature transmitters, indicators or any input
devices that measure thermocouple sensors.
Move the power switch
w
to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select T/C for the FUNCTION, °F or °C for the UNITS, T/C Type
(J, K, E, T, R, S, B, N, L (J-DIN), U (T-DIN), G, C, D or P (Platinel
II)) and internal COLD JUNC ON or OFF (ON is the default).
Connect the PIE 525Plus to the inputs of the device being
calibrated using the proper type of thermocouple wire via the
miniature thermocouple socket.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 0.1°
increments in LOW and 0.01° increments in HIGH resolution by
turning the knob
e
. Press and turn the knob for faster dialing
with 10.0° increments in LOW and 10.00° increments in HIGH
resolution.
Miniature thermocouple connector
Instrument with T/C Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
Thermocouple Wire
HI
TYPE K
22.4°C
43.307mV
°C
1052.50
T/C
+
-
OFF
OVERLOAD
Double Click
Menu
HI
SET
LO
SOURCE
MAX
READ
MIN
READ
Push & Hold
to
Store/Step
Push & Turn
for
Fast Dialing
Diagnostic Thermocouple
RTD & Milliamp Calibrator
2, 3, 4 Wire Detection
Loop Leak Detect ∙ Transmitter Supply
www.
GlobalTestSupply
.com
Find Quality Products Online at: