8
Dos and Don’ts
Never
Never
Never
Never ground any output connector.
Never
Never
Never
Never use any grounded probes at the output.
Do not
Do not
Do not
Do not connect your oscilloscope to the output!
This will immediately destroy the driver and the probe!
For measuring current and voltage you connect the scope to Pin 9 or Pin 4 of the BOB connector
respectively.
Keep connecting cables between power supply and driver as well as the connection between driver
and laser diode as short as possible.
Mount the driver on an appropriate heat sink!
Mount the driver on an appropriate heat sink!
Mount the driver on an appropriate heat sink!
Mount the driver on an appropriate heat sink!
Please
Please
Please
Please be aware that there might be hot surfaces, be careful not to touch them!
be aware that there might be hot surfaces, be careful not to touch them!
be aware that there might be hot surfaces, be careful not to touch them!
be aware that there might be hot surfaces, be careful not to touch them!
Do never connect the oscilloscope to the output connectors!!!!
(Please Note: above picture shows another but similar PicoLAS driver)