5 Installation
Q-521 Miniature Linear Stage
MP145E
Version: 1.0.1
21
Thread-locking adhesive
Mounting the Q-521 on a surface and connecting it to a protective earth conductor
Figure 6: Mounting from above, using a Q-521.2xx as an example
1 Option: Aligning the linear stage (view from below) on the surface with
1.5 m6 x 4 locating pins.
2, 3 Mounting the linear stage on the surface with two M2x6 screws.
1.
Option: Align the Q-521 on the surface with the locating pins.
2.
Mount the Q-521 on the surface.
Maximum torque: 35 Ncm
For mounting from above:
−
If necessary, manually displace the motion platform of the Q-521 to access the
countersunk holes in the base body.
For mounting from below:
−
Maximum screw-in depth in the Q-521: 4 mm
3.
Secure the screws against unintentional loosening, e.g., with thread-locking adhesive.
4.
Check that the Q-521 is affixed firmly.
5.3
Setting Up a Multi-Axis System
The Q-521 can be used in multi-axis systems.
Typical combinations:
XY system (p. 23)
Z system (p. 25) (XZ or XYZ combination)