background image

 

 

 

 

 

 

 

10 

www.phywe.com, © All rights reserved 

 

09057-99 / 2115 

 

 
Fig. 12: Menu structure of the unit – overview 

 

6.5  Control via a PC with “measure” 

As an alternative with regard to the direct control of the unit 
via the control panel, the control of the unit and the represen-
tation of the measurement values can also be realised with 
the aid of a PC and the “measure” software (14414-61). 
 

6.6  Automatic shutdown of high voltage in case of fail-
ure 

In case of interfering signals in the mains supply, the high 
voltage will automatically be cut off in means to protect the 
device from defects (introduced with firmware V

2.1). This 

can be discerned by the fact that the x-ray tube stops to 
gleam. 
Running measurements eventually will not stop during the 
shutdown, so they probably have to be restarted. 
After removing the source of interference, check the door 
status button IV and lock the door, if necessary (LED has to 
be green). Afterwards, the device can be put into operation 
again by pressing the X-ray button V. Thus, the high voltage 
is switched on again and the measurement can be repeated. 
If the device is controlled via a PC, the USB connection has 
to be unplugged and plugged in, anew.  

7 CHECKLIST 

According to this list the device should be checked at least  
two times per year: 
1.  Check if the device shows any damage. 
2.  All glass panes must be undamaged. 
3.  Start the x-ray device according to chapter 6.1, 6.2 of this 

operating instruction. Perform a functional test of the 
status indications, e.g.: 

 

Do all LEDs work porperly? 

  Is symbol „I“ enlighted in operating state 3 (Ta-

ble 3)? 

4.  Check all functionalities according to table 2 of this oper-

ating instruction: 

 

In operating state 1 and 2 it is not possible to acti-
vate the x-radiation.  

 

In operating state 3 the x-radiation is not active, the 
door is locked. 

  Only in operating state 4 the x-radiation is active. 

Does sign 

 appear on the screen (Fig. 11)? 

 

If you activate the x-radiation with the following pa-
rameters: anode current 1

mA, anode voltage 

35

kV, the x-ray tube should gleam brightly. 

 
The device is only ready for use if the tests are performed in 
the right way and successfully. Even if 

only one

 test fails the 

device must not be used. Please contact the Phywe service 
in this case. ([email protected]). 

8 TECHNICAL 

DATA 

 

Operating temperature range: 5–40

°C, typically 25

°C 

 

Rel. humidity < 70

 

Microprocessor-controlled XR 4.0 X-ray expert unit with a 
central safety monitoring system and two independent 
monitoring circuits for the door position and two inde-
pendent monitoring circuits for the actuator of the door 
locking system 

  4 X-ray tubes that are visible during the operation (Fe, 

Cu, Mo, and W) 

  Lead enforced acrylic class windows, safe according to 

DIN EN 61010 

  Integrated display unit for displaying measurement val-

ues and device parameters 

  “Display Connect” interface for the connection of a large-

scale display unit 

  Experiment chamber accessible during the operation via 

a working channel 

  Integrated LED line for interior lighting (can be activated 

as required) 

 

Internal and external socket panel for easy cabling 

 

Loudspeaker for the acoustic indication of measurements 
with the Geiger-Müller counter tube 

 

N

2

max 5 bar 

 

Lockable tray for storing accessories 

 

High voltage: 5.0–35.0

kV 

 

Emission current: 0.0–1.0

mA 

 

counter tube voltage: 100–600

 Counting 

time: 

0.5–100

 

Exposure time: 0–100 minutes 

Summary of Contents for XR 4.0

Page 1: ...se in dry and dust free rooms where there is no risk of explosion Prior to connecting the unit to the mains power supply ensure that the protective conductor of the power supply unit is correctly conn...

Page 2: ...g mobile phones should not be used in the direct vicinity of the unit The connected cables must not be longer than 2 m Electrostatic charges or similar electromagnetic phe nomena HF bursts indirect li...

Page 3: ...ography of objects and observation with the aid of a fluoroscopic screen Preparation of X ray images of irradiated objects Detection of the ionising effect of X rays dosimetry Detection of Bragg refle...

Page 4: ...ww phywe com All rights reserved 09057 99 2115 Fig 2 Front view of the XR 4 0 X ray expert unit 09057 99 including X ray tube Fig 3 View from the right side XR 4 0 X ray expert unit 09057 99 1 4 2 3 5...

Page 5: ...side of the unit looping through This enables for ex ample the control or read out from the outside of a digital camera that is located inside the experiment chamber 5 2 4 Working channel D The worki...

Page 6: ...sible to swith on the unit Left LED green X ray can be activated right LED green X ray is activated VI Button no name The selection of the menu function will be displayed on the screen above on the bo...

Page 7: ...4 0 X ray expert unit Make sure that the plug in is inserted completely Otherwise it is not going to work Handle the plug ins with care 5 6 Socket panel inside the experiment chamber 5 Fig 7 Socket pa...

Page 8: ...g door is not locked with the lock bar If so unlock the door using but ton IV Fig 5 switch on the device to use the button The door should not be open either fix it with the S LOCK 6 2 Start up Connec...

Page 9: ...and 35 0 kV and the emission current to a value between 0 00 mA and 1 00 mA 6 4 2 Goniometer Menu and parameterisation of the goniometer 09057 10 The operation of the goniometer in the X ray unit is...

Page 10: ...all functionalities according to table 2 of this oper ating instruction In operating state 1 and 2 it is not possible to acti vate the x radiation In operating state 3 the x radiation is not active t...

Page 11: ...R 4 0 X ray Chemical set for edge absorption 09058 11 XR 4 0 X ray Crystal holder for Laue pattern 09058 09 XR 4 0 X ray holder for powder probes diffrac tometry 09057 08 XR 4 0 X ray film holder 0905...

Page 12: ...PHYWE Systeme GmbH Co KG Customer Service Robert Bosch Breite 10 37079 G ttingen Germany Telephone 49 0 551 604 274 Fax 49 0 551 604 246 13 APPENDIX Symbols Warning Warning radioactive sub stance or...

Page 13: ...e mount for optical bench h 30 mm x 10 09058 04 XR 4 0 X ray Compton attachment for x ray unit 11 08286 01 Slide mount for optical bench h 30 mm x 12 09057 01 XR 4 0 X ray Diaphragm tube d 1 mm 09057...

Reviews: