The unit has four device statuses that are listed in Table 2. 
Depending on the status, the unit can be controlled entirely 
via the buttons on the control panel together with the graphi-
cal representation. 

 

Fig. 10: Screen for controlling the unit at the front of the unit. 
The x-radiation is not activated. 

 

Fig. 11: Screen for controlling the unit at the front of the unit. 
The x-radiation is active. Sign 

 appears on the screen. 

 
The change of the representation and the control of the unit 
via the menu are performed via the buttons VI–X (Fig. 5). 
As soon as the x-radiation is activated a yellow symbol ap-
pears on the screen (See Fig. 11). 
 

6.4 Menu 

structure 

On the main level, the menu structure of the unit includes the 
following items: 

1. X-ray 

parameters 

2. Goniometer 
3. Timer 
4. Settings 
5. Info 
6. Large 

Display 

The buttons VI–X on the control panel are used to navigate 
through the menu, to change the settings, and to quit the 
menu. 
 

6.4.1 X-ray parameters 

This menu is used to set the high voltage and the beam cur-
rent. By selecting the submenus for the high voltage or beam 
current and by pressing the dynamic buttons on the right of 
the display unit, the high voltage of the tube can be set to a 
value between 0.0 kV and 35.0 kV and the emission current 
to a value between 0.00 mA and 1.00 mA. 
 

6.4.2 Goniometer 

Menu and parameterisation of the goniometer (09057-10). 
The operation of the goniometer in the X-ray unit is described 
in the operating instructions for 09057-10. 
 

6.4.3 Rotation stage 

Menu and parameterisation of the rotation stage XRstage 
(09057-42). The operation of the rotation stage inside of the 
X-ray unit is described un the operating instructions for 
09057-42. 
 

6.4.4 Timer 

This menu is used to define the start and stop conditions for 
the experiments (X-radiation ON/OFF). 
 

6.4.5 Settings 

This menu is used to define certain fundamental settings. 
 

6.4.6 Info 

Menu for status information concerning the unit. 
 

6.4.7 Large Display 

If a sender (09057-09) is connected  to the XR 4.0 expert unit 
and transfers date to a large display in this part of the menu 
the parameter for the display can be set. 

Table 2: Overview concerning the three possible operating states 

 

Status/X-radiation 

Description 

Button IV 

Button V 

Sign  

panel 

(Fig. 5, I) 

Sign 

Display

(Fig. 11) 

1  Cannot be switched on 

The sliding door is not 

closed and not locked. 

white white 

off  off 

2  Cannot be switched on 

The sliding door is closed 

but not locked 

Left LED 

green 

white off 

off 

3  Can be switched on 

The unit is in a safe state 

and the X-radiation can 
be activated. 

Right LED 
green 

Left LED 
green 

on off 

4 On 

The unit is in a safe state 

and the X-radiation is ac-
tive. 

Right LED 
green 

Right LED 
green 

on on 

Summary of Contents for XR 4.0

Page 1: ...se in dry and dust free rooms where there is no risk of explosion Prior to connecting the unit to the mains power supply ensure that the protective conductor of the power supply unit is correctly conn...

Page 2: ...g mobile phones should not be used in the direct vicinity of the unit The connected cables must not be longer than 2 m Electrostatic charges or similar electromagnetic phe nomena HF bursts indirect li...

Page 3: ...ography of objects and observation with the aid of a fluoroscopic screen Preparation of X ray images of irradiated objects Detection of the ionising effect of X rays dosimetry Detection of Bragg refle...

Page 4: ...ww phywe com All rights reserved 09057 99 2115 Fig 2 Front view of the XR 4 0 X ray expert unit 09057 99 including X ray tube Fig 3 View from the right side XR 4 0 X ray expert unit 09057 99 1 4 2 3 5...

Page 5: ...side of the unit looping through This enables for ex ample the control or read out from the outside of a digital camera that is located inside the experiment chamber 5 2 4 Working channel D The worki...

Page 6: ...sible to swith on the unit Left LED green X ray can be activated right LED green X ray is activated VI Button no name The selection of the menu function will be displayed on the screen above on the bo...

Page 7: ...4 0 X ray expert unit Make sure that the plug in is inserted completely Otherwise it is not going to work Handle the plug ins with care 5 6 Socket panel inside the experiment chamber 5 Fig 7 Socket pa...

Page 8: ...g door is not locked with the lock bar If so unlock the door using but ton IV Fig 5 switch on the device to use the button The door should not be open either fix it with the S LOCK 6 2 Start up Connec...

Page 9: ...and 35 0 kV and the emission current to a value between 0 00 mA and 1 00 mA 6 4 2 Goniometer Menu and parameterisation of the goniometer 09057 10 The operation of the goniometer in the X ray unit is...

Page 10: ...all functionalities according to table 2 of this oper ating instruction In operating state 1 and 2 it is not possible to acti vate the x radiation In operating state 3 the x radiation is not active t...

Page 11: ...R 4 0 X ray Chemical set for edge absorption 09058 11 XR 4 0 X ray Crystal holder for Laue pattern 09058 09 XR 4 0 X ray holder for powder probes diffrac tometry 09057 08 XR 4 0 X ray film holder 0905...

Page 12: ...PHYWE Systeme GmbH Co KG Customer Service Robert Bosch Breite 10 37079 G ttingen Germany Telephone 49 0 551 604 274 Fax 49 0 551 604 246 13 APPENDIX Symbols Warning Warning radioactive sub stance or...

Page 13: ...e mount for optical bench h 30 mm x 10 09058 04 XR 4 0 X ray Compton attachment for x ray unit 11 08286 01 Slide mount for optical bench h 30 mm x 12 09057 01 XR 4 0 X ray Diaphragm tube d 1 mm 09057...

Reviews: