background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

09057-99 / 2115

 

5.7  External socket panel on the right-hand side (

6

The external socket panel (Fig 8) on the right-hand side of 
the unit is the counterpart of the socket panel inside the ex-
periment chamber. 

 

Fig. 8: External socket panel on the right-hand side of the unit 
 
Here, the following components can be connected. The 
names that are stated on the unit are printed in italics. 
 
Lower line of the socket panel: 

 

PC:

 For the connection to the control PC via the “meas-

ure” USB 2.0 port. 

 

Service

: For the connection of a specially configured PC 

for updating the device settings and for diagnosis pur-
poses (for authorised specialist personnel only). 

 

Display:

 For the connection of a “Display Connect” RF 

adapter (09057-19). It is used to display measurement 
values and device parameters on the large-scale display 
unit (07157-93). 

 

USB 2.0:

 Socket for connecting digital cameras etc. (di-

rect connection to the socket panel inside the experiment 
chamber). 

 

N

2

:

 For feeding in protective gas or for the connection to 

a vacuum pump (direct connection to the socket panel 
inside the experiment chamber). 

 
Upper line of the socket panel: 

 

X RED:

 For connecting the X-ray energy detector 09058-

30 to the multi-channel analyser 13727-99; BNC signal 
cable, supply cable (direct connection to the socket panel 
inside the experiment chamber). 

 

Aux:

 Multi-pole socket for connecting various devices in 

the experiment chamber (direct connection to the socket 
panel inside the experiment chamber). 

 

Max 600 V

: 2 x 4-mm-sockets, e.g. for charging the ca-

pacitor plates (order no. 09057-05) for dosimetry experi-
ments (direct connection to the socket panel inside the 
experiment chamber). 

6 OPERATION 

This section describes the start-up of the unit and provides 
an overview of its operation. Please read this section care-
fully in order to avoid problems or malfunctions. 
 

6.1 Transport 

  Do not stress the drawer during transport. The device 

should stand only on its feet.  

  During transport make sure that the sliding door is not 

locked with the lock bar. If so, unlock the door using but-
ton IV, Fig 5 (switch on the device to use the button). The 
door should not be open either – fix it with the S-LOCK. 

 

6.2 Start-up 

Connect the unit to the power supply via the supplied power 
cable with an IEC connector. The socket for this purpose is 
located at the back of the unit (see Fig. 9). 

 

Fig. 9: Back of the unit with the socket for the power cable 
and the central ON/OFF switch 
 

6.3  Starting the unit 

The central ON/OFF switch of the unit is located at the back 
of the unit (Fig. 9). Actuate this switch to switch the unit on. 
After the start, the unit will perform an automatic self-test 
(safety test). For this purpose, the operator must open the 
sliding door completely one time and then close it again. The 
device status that is thus determined will be indicated on the 
display unit in the control panel via the colours of the but-
tons IV and V. 
In order to open the sliding door, push it first against the stop 
on the right-hand side.The S-Lock is now unlocked and the 
door can be opened. 
If the unit had been switched off with the door being locked, 
the door will be automatically unlocked once the unit is 
switched on. The operator must then open the radiation pro-
tection door once and close it again (check of the safety cir-
cuit; described in the operating instructions and on the dis-
play unit). 
If the X-ray plug-in is completely inserted and the door is 
closed with the use of the SLOCK the left LED of button IV is 
glowing in green. All other LEDs are enlighted in white. Now, 
the unit is ready for use.  
If this is not the case, contact the Service of PHYWE for help. 

Caution:  

  It is only possible to lock the sliding door, if the 

plug in with the X-ray tube is completely inserted 
into the device. 

 

if you switch off the device while it is locked you 
have to switch it on again to unlock the door. 

 

When used for the first time, the X-ray tubes should not run 
at full power. Instead, we recommend letting the tubes run for 
approximately 10 minutes at the maximum beam current, but 
with an acceleration voltage that is limited to 25

kV maximum. 

This procedure must also be repeated if a tube has not been 
used for several weeks. 
 
 

 

Summary of Contents for XR 4.0

Page 1: ...se in dry and dust free rooms where there is no risk of explosion Prior to connecting the unit to the mains power supply ensure that the protective conductor of the power supply unit is correctly conn...

Page 2: ...g mobile phones should not be used in the direct vicinity of the unit The connected cables must not be longer than 2 m Electrostatic charges or similar electromagnetic phe nomena HF bursts indirect li...

Page 3: ...ography of objects and observation with the aid of a fluoroscopic screen Preparation of X ray images of irradiated objects Detection of the ionising effect of X rays dosimetry Detection of Bragg refle...

Page 4: ...ww phywe com All rights reserved 09057 99 2115 Fig 2 Front view of the XR 4 0 X ray expert unit 09057 99 including X ray tube Fig 3 View from the right side XR 4 0 X ray expert unit 09057 99 1 4 2 3 5...

Page 5: ...side of the unit looping through This enables for ex ample the control or read out from the outside of a digital camera that is located inside the experiment chamber 5 2 4 Working channel D The worki...

Page 6: ...sible to swith on the unit Left LED green X ray can be activated right LED green X ray is activated VI Button no name The selection of the menu function will be displayed on the screen above on the bo...

Page 7: ...4 0 X ray expert unit Make sure that the plug in is inserted completely Otherwise it is not going to work Handle the plug ins with care 5 6 Socket panel inside the experiment chamber 5 Fig 7 Socket pa...

Page 8: ...g door is not locked with the lock bar If so unlock the door using but ton IV Fig 5 switch on the device to use the button The door should not be open either fix it with the S LOCK 6 2 Start up Connec...

Page 9: ...and 35 0 kV and the emission current to a value between 0 00 mA and 1 00 mA 6 4 2 Goniometer Menu and parameterisation of the goniometer 09057 10 The operation of the goniometer in the X ray unit is...

Page 10: ...all functionalities according to table 2 of this oper ating instruction In operating state 1 and 2 it is not possible to acti vate the x radiation In operating state 3 the x radiation is not active t...

Page 11: ...R 4 0 X ray Chemical set for edge absorption 09058 11 XR 4 0 X ray Crystal holder for Laue pattern 09058 09 XR 4 0 X ray holder for powder probes diffrac tometry 09057 08 XR 4 0 X ray film holder 0905...

Page 12: ...PHYWE Systeme GmbH Co KG Customer Service Robert Bosch Breite 10 37079 G ttingen Germany Telephone 49 0 551 604 274 Fax 49 0 551 604 246 13 APPENDIX Symbols Warning Warning radioactive sub stance or...

Page 13: ...e mount for optical bench h 30 mm x 10 09058 04 XR 4 0 X ray Compton attachment for x ray unit 11 08286 01 Slide mount for optical bench h 30 mm x 12 09057 01 XR 4 0 X ray Diaphragm tube d 1 mm 09057...

Reviews: