background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

09057-99 / 2115

 

  Do not start the unit if the mains power cable or the unit 

itself are damaged. 

  Switch the unit off after using it. Continuous operation is 

not permissible. The maximum runtime per day is 
10 hours. It is not a syfety risk if you the tube for longer 
intervals but this could reduce the lifetime of the tube.  

  Start the unit at least twice per year and check its safety 

functions. The tests that are included in the checklist at 
the end of this document must be performed. 
If damaged, the unit must be returned to the manufac-
turer for repair. Only the manufacturer or a certified com-
pany that has been appointed by the manufacturer are 
authorised to repair and maintain the unit. 
 

 

  Since X-ray units generate dangerous radiation that is 

hazardous to health, only trained and qualified persons 
are authorised to start the X-ray unit in accordance with 
the local, country-specific rules and regulations. 

  In Germany: When working with the X-ray unit, the man-

datory measures and duties that are outlined in the Ger-
man X-ray ordinance (Röntgenverordnung, RÖV) must 
be strictly complied with. 

 

The requirements of the certificate are to be followed 

 

In particular, the operator must ensure that 

o

 

the X-ray unit is protected against access by unau-
thorised persons; 

o

 

the unit is not in use longer than necessary; 

o

 

persons working with the unit do not remain in the 
direct vicinity of the unit longer than absolutely 
necessary. 

 

The use of the unit is prohibited if 

o

 

the sliding door that is made of lead-containing 
acrylic glass and that is used for opening the ex-
periment chamber, or the other protective glass 
windows that are used to observe the X-ray unit 
and the experiment chamber, are damaged; 

o

 

the fan at the X-ray plug-in side inside the unit is 
inoperative (acoustic check); 

o

 

the safety circuits for interrupting the X-ray opera-

tion when the sliding door is opened do not oper-
ate properly. 

 

How to clean: 

o

 

Do not use diluter/thinner! 

o

 

Clean with cloth and appropriate cleaning agent. 

 

Transport 

o

 

Do not stress the drawer during transport. The 
device should stand only on its feet.  

o

 

During transport make sure that the sliding 
door is not locked with the lock bar. If so, 
unlock the door using button IV, Fig 5 (switch 
on the device to use the button). The door 
should not be open either – fix it with the 
S-LOCK. 

2 OPERATING 

INSTRUCTIONS 

  The device fulfils the technical requirements that are 

summarised in the current guidelines of the European 
Community. The characteristics of the product entitle it to 
bear the CE mark. 

 

The unit must be used under the supervision of an expert 
and in the electromagnetically controlled environment at 

research, teaching, and training facilities (schools, uni-
versities, institutes, and laboratories). 

 

This means that, in such an environment, radio transmis-
sion devices, e.g. mobile phones, should not be used in 
the direct vicinity of the unit. The connected cables must 
not be longer than 2

m. 

  Electrostatic charges or similar electromagnetic phe-

nomena (HF, bursts, indirect lightning discharge, etc.) 
may affect the unit so that it will not work within the 
specified data range. 
Occuring interfering signals may result in an automatic 
shutdown of the high voltage during measurement! In 
this case please follow section 6.6. 
The following measures reduce or eliminate potential in-
terferences: avoid carpets; provide equipotential bond-
ing; perform the experiments on a conductive, earthed 
surface; use shields and shielded cables. Do not use ra-
diofrequency transmitters (radio sets, mobile phones) in 
the direct vicinity of the unit. After a total exit by actuating 
the mains power switch, perform a reset. 

This unit corresponds to class A of the standard DIN 
EN 61326 and may be used without any restriction only in 
non-residential areas. If – although the use of the unit is re-
stricted to special classrooms in a school or another training 
facility – electromagnetic interferences occur in the surround-
ing residential area, the operator can be required to take the 
appropriate measures (e.g. shielding, long distance with re-
gard to sensitive equipment, short periods of use, use of the 
shortest possible connecting cables, etc.) and to pay for 
these measures. 

3  INSTRUCTIONS OF THE GERMAN X-RAY 
ORDINANCE 

 

In Germany, the operation of the X-ray unit is subject to the 
regulations that are outlined in the X-ray ordinance (Rönt-
genverordnung, RöV). In accordance with this ordinance, the 
unit fulfils the mandatory requirements of an X-ray unit for 
schools as well as those of a full-protection device. 
In Germany, the operation of the unit is not subject to ap-
proval, but to reporting. Thus, report the unit to the appropri-
ate controlling institution. 
For reporting, the test certificate and a copy of the design ap-
proval must be submitted to the responsible authorities. Only 
specifically trained and instructed personnel is authorised to 
operate the unit. When this unit is used as an x-ray device for 
schools a radiation protection officer must be authorized. 
If the unit operates at the maximum operating data, the local 
dose rate at a distance of 0.1

m from the parts of the housing 

that can be touched is less than 1

 μ

Sv/h.  

Two independent safety circuits monitor the opening of the 
sliding door that leads to the experiment chamber. 
X-rays can only be generated when the sliding door is prop-
erly locked. The safety circuits also prevent the maximum 
permissible operating values of the tube from being ex-
ceeded. 
Do not use the device if the X-ray unit is manipulated in a 
way that is not necessary for its set-up or for performing ex-
periments. It is absolutely forbidden to loosen the safety 
screws of the sheet steel housing or to tamper with the win-
dows. Only the manufacturer is authorised to repair the unit. 
For any operation of the unit outside of Germany, the appli-
cable local rules and regulations must be complied with. 

Summary of Contents for XR 4.0

Page 1: ...se in dry and dust free rooms where there is no risk of explosion Prior to connecting the unit to the mains power supply ensure that the protective conductor of the power supply unit is correctly conn...

Page 2: ...g mobile phones should not be used in the direct vicinity of the unit The connected cables must not be longer than 2 m Electrostatic charges or similar electromagnetic phe nomena HF bursts indirect li...

Page 3: ...ography of objects and observation with the aid of a fluoroscopic screen Preparation of X ray images of irradiated objects Detection of the ionising effect of X rays dosimetry Detection of Bragg refle...

Page 4: ...ww phywe com All rights reserved 09057 99 2115 Fig 2 Front view of the XR 4 0 X ray expert unit 09057 99 including X ray tube Fig 3 View from the right side XR 4 0 X ray expert unit 09057 99 1 4 2 3 5...

Page 5: ...side of the unit looping through This enables for ex ample the control or read out from the outside of a digital camera that is located inside the experiment chamber 5 2 4 Working channel D The worki...

Page 6: ...sible to swith on the unit Left LED green X ray can be activated right LED green X ray is activated VI Button no name The selection of the menu function will be displayed on the screen above on the bo...

Page 7: ...4 0 X ray expert unit Make sure that the plug in is inserted completely Otherwise it is not going to work Handle the plug ins with care 5 6 Socket panel inside the experiment chamber 5 Fig 7 Socket pa...

Page 8: ...g door is not locked with the lock bar If so unlock the door using but ton IV Fig 5 switch on the device to use the button The door should not be open either fix it with the S LOCK 6 2 Start up Connec...

Page 9: ...and 35 0 kV and the emission current to a value between 0 00 mA and 1 00 mA 6 4 2 Goniometer Menu and parameterisation of the goniometer 09057 10 The operation of the goniometer in the X ray unit is...

Page 10: ...all functionalities according to table 2 of this oper ating instruction In operating state 1 and 2 it is not possible to acti vate the x radiation In operating state 3 the x radiation is not active t...

Page 11: ...R 4 0 X ray Chemical set for edge absorption 09058 11 XR 4 0 X ray Crystal holder for Laue pattern 09058 09 XR 4 0 X ray holder for powder probes diffrac tometry 09057 08 XR 4 0 X ray film holder 0905...

Page 12: ...PHYWE Systeme GmbH Co KG Customer Service Robert Bosch Breite 10 37079 G ttingen Germany Telephone 49 0 551 604 274 Fax 49 0 551 604 246 13 APPENDIX Symbols Warning Warning radioactive sub stance or...

Page 13: ...e mount for optical bench h 30 mm x 10 09058 04 XR 4 0 X ray Compton attachment for x ray unit 11 08286 01 Slide mount for optical bench h 30 mm x 12 09057 01 XR 4 0 X ray Diaphragm tube d 1 mm 09057...

Reviews: