Notes on operation
PHYSIOTHERM-S
9
4.4
Instrument Errors
If a functional error is detected during the automatic self-test routine or during
operation, a corresponding note will be displayed on the LCD. A numeric error code
will be shown, e.g. E02. Both operator errors and internal errors in the unit are
displayed as an error code with text information.
4.4.1
Operator errors
The following operator errors might be displayed:
E01:
Please check electrodes and cables
Cause:
An optimum tuning point could not be found; therefore, the applicators,
their ESD and the correct connection to the electrode cable should be
checked again.
E02:
Matching...Please wait!
Cause:
For reasons of safety, the effective output must not be set higher than
100 W, as long as the tuning process is still in progress. The reason for
this is that if tuning takes place at a higher effective output, heat can be
clearly felt.
4.4.2
Internal errors
With the exception of temperature errors, the treatment must be stopped
immediately in case of internal errors in the unit. The unit can only be switched off
and on again by means of the
power switch <1>
. If the error message appears
again, make a note of the error code, switch the unit off again and notify your
authorized service partner.
In case of display failure or other obvious defects, switch the unit off immediately by
means of the
power switch <1>
and notify your authorized service partner!