Notes on operation
8
PHYSIOTHERM-S
4
Notes on operation
4.1
Preparing the unit
The unit has been completely assembled in the factory and is ready for use except
for connection of the electrodes.
Proceed as follows in order to prepare the unit for operation:
(1)
Make sure that the voltage rating on the rating plate conforms to the system
voltage of the building.
(2)
Insert the required electrodes into the recesses at the end of the electrode arms
and fasten the electrodes with the locking screws.
(3)
Plug the electrode connection cable into the socket on the back of the unit.
Attach the connecting cable in the cable holders on the electrode arms.
(4)
Set the power switch to the off position.
(5)
Plug the high-load connector for non-heating appliances into the corresponding
socket on the back of the unit.
4.2
Check for operational safety
For your personal safety you should:
•
Check the condition of the housing and the insulation of electrodes, electrode
connection cable and the power supply cable. Also make sure that the cables
have been routed correctly.
•
Insert the power plug into an earthed (!) socket outlet.
Warning
The unit and the electrodes must be positioned so that there is no
danger of personal injury! Therefore, you must read and observe the
safety instructions in section
on page 46 and the
on page 2 before putting the unit into operation!
4.3
Setting the Line Voltage
The instrument is equipped with a line voltage selector on
the rear side (next to the line fuses). Use this selector to
switch the line voltage of the instrument to 230 VAC or 115
VAC. It is not necessary to change the line fuses T16A.