Basic
settings
PHYSIOTHERM-S
45
9
Basic settings
In the basic settings, you can set different parameters that are rarely altered.
Unless you alter them, the indicated parameters are set to the following default
values:
Parameter Default
value Range
Pulse duration
400
μ
s
200 - 600
μ
s (default: 400
μ
s)
Volume tone
50 %
0 – 100 %
Service code
0
LCD language
any
DE (German), EN
(
English), ES (Spanish),
FR (
French), IT (Italian), PT (Portuguese), RU
(Russian), TR (Turkish)
LCD brightness
70 %
10 - 100 %
LCD contrast
70 %
0 - 100 %
9.1
Changing the basic settings
To alter a parameter, proceed as follows:
(1)
Switch off the instrument.
(2)
Press the
data selector <2>
and hold it down.
(3)
Switch on the instrument with the
power switch <1>
and wait (with the data
selector depressed) until the basic settings menu is displayed.
(4)
Turn the data selector until the desired parameter line is selected.
(5)
Press the data selector. The value to be changed in this line is selected.
(6)
Set the desired parameter value by turning the data selector.
(7)
Confirm the set value by pressing the data selector. The entire line is then
selected again.