5.5.3
Storing the calibration in permanent memory
After running the calibration procedure (see Section 5.5.2) the calibration values are stored in
RAM. When the camera is turned off, their values are lost.
To prevent this, the calibration values must be stored in flash memory. This can be done by
clicking on the property
BadPixelCorrection_SaveToFlash
(in category
BadPixelCorrection
). Wait
until the command has been finished.
Storing the calibration in permanent memory overwrites the factory calibration.
5.6
Column FPN Correction
Due to the readout structure of the image sensors there is a column-wise fixed pattern noise
(FPN). The Column FPN Correction (ColCorrection) adds or subtracts a fixed value to a column.
These values are obtained by a calibration process. The ColCorrection of the camera was
calibrated at Photonfocus production facility.
5.6.1
Enable / Disable the Column FPN Correction
The Column FPN Correction can be enabled or disabled with the property
ColCorrection_Enable
(in category
Correction
/
ColCorrection
). By default the correction is enabled.
5.6.2
Calibration of the Column FPN Correction
The Column FPN Correction of the camera is correctly calibrated at Photonfocus production
facility. Although a new calibration is normally not required, you can recalibrate the Column
FPN Correction with the following instructions:
1.
Setup the camera to the mode where it will be usually used. (Exposure time, ...). The
width and height should be set to its maximal value.
✎
If different exposure times will be used, calibrate the camera under the longest
exposure time.
2.
Put the camera in free-running mode by setting the property
TriggerMode
to
Off
.
3.
Start grabbing of the camera by clicking on the
Play
button.
4.
Wait until the camera has achieved working temperature.
5.
Close the lens of the camera or put a cap on the lens. The calibration requires an uniform
dark image. The black level offset should be set so that all pixels (except defect pixels)
have values above 0.
6.
Set the visibility of the Device Control window of the PF_GEVPlayer to
Expert
.
7.
Run the command
ColCorrection_CalibrateBlack
(category
ColCorrection
) by clicking on
the corresponding button. The camera transmits a test image during calibration.
8.
Run the command
ColCorrection_Update
by clicking on the corresponding button. Read the
ColCorrection_Busy
value which should be 0 after the calibration has finished. Repeat this
step if its value is not 0. If the
ColCorrection_Busy
value doesn’t show 0 after various tries,
check if the camera receive triggers or set the
TriggerMode
of the camera to
Off
.
5.6 Column FPN Correction
59
Summary of Contents for DR1-D3360-192-G2-8
Page 1: ...MAN071 03 2015 V1 0 Photonfocus MV1 D3360 G2 Camera Series CMOS camera with GigE interface...
Page 10: ...3 How to get started GigE G2 Figure 3 3 PFInstaller components choice 10...
Page 22: ...4 Product Speci cation Figure 4 1 Photonfocus D3360 GigE camera series with C mount lens 22...
Page 76: ...5 Functionality 76...
Page 92: ...6 Hardware Interface 92...
Page 112: ...9 Warranty 112...
Page 114: ...10 References 114...
Page 119: ...C Document Revision History Revision Date Changes 1 0 March 2016 First version 119...