Te c h n i c a l S p e c i f i c a t i o n s
B-5
P
h
ili
ps
Me
dic
a
l
Sy
s
tem
s
B
Display
Controls and indicators
Pacemaker Detection
On detection of a pacemaker (in advanced mode or with M3873A/M3874A FR2+
ECG assessment module), provides screen display of
PACEMAKER
DETECTED
alert. M3860A includes pacemaker artifact in ECG display. In both models,
pacemaker artifact is removed from the signal for rhythm analysis.
category
nominal specifications
category
nominal specifications
Monitored ECG Lead
ECG information is received from adult defibrillator pads in anterior-anterior (Lead
II) position or from FR2 infant/child reduced-energy defibrillator pads in
anterior-posterior position. (Displayed on M3860A only.)
ECG information can also be displayed in the M3860A using the FR2+ ECG
assessment module.
N O T E : The ECG display provided by the FR2+ Defibrillator is not is not
intended to provide diagnostic or ST segment interpretation.
Display Range
(M3860A only)
Differential: ±2 mV full scale, nominal.
Screen Type
High-resolution liquid crystal display (LCD) with backlight.
Screen Dimensions
2.8" wide x 2.3" high (70 mm x 58 mm).
Sweep Speed
(M3860A only)
23 mm/s nominal.
ECG Display
3 second-segments displayed (M3860A only).
Frequency Response
(Bandwidth)
Nondiagnostic rhythm monitor 1 Hz to 20 Hz (-3 dB), nominal.
Sensitivity
1.16 cm/mV, nominal.
Heart Rate Displayed
(Normal Sinus Rhythm)
30 to 300 bpm, updated each analysis period. Displayed (M3860A only) during
monitoring and advanced modes.
category
nominal specifications
LCD Screen
High-resolution, backlighted LCD screen, displays text messages and (model
M3860A only) ECG.
Summary of Contents for HeartStart FR2+
Page 4: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...
Page 10: ...M3860A M3861A HEARTSTART FR2 P h i l i p s M e d i c a l S y s t e m s Notes ...
Page 14: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 20: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 42: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 66: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 78: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 94: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 101: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...