Maintaining, Testing, and Troubleshooting Your HeartStart FR2+
4-13
P
h
ili
ps
Me
dic
a
l
Sy
s
tem
s
4
Voice prompts:
ANALYZING
INTERRUPTED
or
CANNOT
ANALYZE
or
STOP
ALL
MOTION
•
The patient is being moved or jostled.
•
Radio or electrical sources are
interfering with ECG analysis.
•
The environment is dry and movement
around the patient is causing static
electricity to interfere with ECG
analysis.
•
Stop CPR; do not touch the
patient. Minimize patient motion. If
the patient is being transported,
stop the vehicle if needed.
•
Check for possible causes of
radio and electrical interference
and remove them from the area.
•
Responders and bystanders
should minimize motion,
particularly in dry environments
that can generate static electricity.
Voice and screen prompts:
NO
SHOCK
DELIVERED
The patient impedance is not appropriate
for the FR2+ to deliver a shock.
•
Make sure the defibrillator pads
are correctly positioned on the
patient according to the diagram
on the back of the pads.
•
Make sure the defibrillator pads
connector is completely inserted
in the connector socket.
•
Press the defibrillator pads firmly
to the patient's chest.
•
Replace the defibrillator pads if
necessary.
Voice prompt:
SHOCK
BUTTON
NOT
PRESSED
Shock has been advised but not delivered
within 30 seconds. (FR2+ has been
disarmed.)
•
When next prompted, press the
Shock button to deliver shock.
symptom
possible cause
recommended action
Summary of Contents for HeartStart FR2+
Page 4: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...
Page 10: ...M3860A M3861A HEARTSTART FR2 P h i l i p s M e d i c a l S y s t e m s Notes ...
Page 14: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 20: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 42: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 66: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 78: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 94: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 101: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...