6-1
P
h
ili
ps
Me
dic
a
l
Sy
s
tem
s
6
6
Setup and Advanced Mode Features
Setup Overview
The “setup” of the HeartStart FR2+ Defibrillator is made up of several
programmable aspects, or parameters, of FR2+ operation. Some setup
parameters govern specific features that are not related to the patient care
protocol, some are used to define the automatic patient care protocol used
by the FR2+, and some provide options for manual override of the protocol
during use.
The FR2+ comes with a factory default setup designed to meet the needs of
most users. If desired, your Medical Director can revise the setup. Even if no
changes are made, however, it is a good idea to understand the setup of your
FR2+ and how the different parameter settings affect the way the device
works.
Non-protocol parameters
The parameters listed in the following table enable features of FR2+
operation that are not related to the patient treatment protocol. The table
describes each of these non-protocol parameters, lists the settings available
for it, and identifies the default setting.
parameter
settings
default
description
speaker volume
1, 2, 3, 4,
5, 6, 7, 8
8
Sets volume of the FR2+’s speaker. 1 is lowest; 8 highest.
The speaker is used for voice prompts and the
charge-done tone.
record voice
YES, NO
NO
Enables or disables the audio recording during incident.
Voice recording requires use of a data card.
ECG display
(M3860A only)
ON, OFF
ON
Enables (ON) or disables (OFF) ECG display on the
screen.
FR2+ rhythm analysis does not require ECG display to be
on.
(ECG display cannot be changed from default OFF
for M3861A.)
autosend PST
ON, OFF
OFF
Enables (ON) or disables (OFF) transmission of the
results of the FR2+’s periodic selftests (PST) from its
infrared communications port.
Summary of Contents for HeartStart FR2+
Page 4: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...
Page 10: ...M3860A M3861A HEARTSTART FR2 P h i l i p s M e d i c a l S y s t e m s Notes ...
Page 14: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 20: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 42: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 66: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 78: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 94: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 101: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...