Defibrillation warnings and cautions
CAUTION
Use only Philips-approved data cards. The HeartStart FR2+ may perform
improperly if non-approved accessories are used.
safety level
possible electrical interference with ECG monitoring
WARNING
Radio-frequency (RF) interference from devices such as cellular
phones and two-way radios can cause improper HeartStart FR2+
operation. The HeartStart FR2+ should be used at least 6 feet (2
meters) away from RF devices, as stated in accordance with EN
61000-4-3:1996.
safety level
possible improper device performance
safety level
possible shock hazard
WARNING
Defibrillation current can cause operator or bystander injury. Do not
touch the patient during defibrillation. Disconnect the pads connector
from the HeartStart FR2+ before using any other defibrillator.
safety levels
possible ECG misinterpretation
WARNING
For safety reasons, some very low-amplitude or low-frequency heart
rhythms may not be interpreted by the HeartStart FR2+ as shockable
VF rhythms. Also, some VT rhythms may not be interpreted as
shockable rhythms.
WARNING
Handling or transporting the patient during heart rhythm analysis can
cause an incorrect or delayed diagnosis. If the HeartStart FR2+ gives
a
SHOCK
ADVISED
prompt, keep the patient as still as possible for at
least 15 seconds so the HeartStart FR2+ can reconfirm the rhythm
analysis before a shock is delivered.
Summary of Contents for HeartStart FR2+
Page 4: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...
Page 10: ...M3860A M3861A HEARTSTART FR2 P h i l i p s M e d i c a l S y s t e m s Notes ...
Page 14: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 20: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 42: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 66: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 78: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 94: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Page 101: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...