Customer Engineer Manual
Page 2-38
S8
Not used Pause Duration for Dial Delay. Not
supported. Default is ASCII 1
S9
Not used Reaction Time to Received Carrier.
Default is ASCII 6
S10
Not used Time Delay from Loss of Carrier to
Clearing of Connection. The synchronisation
may be lost for up to 3 seconds.
Default is ASCII 7
S11
Not used Duration and Interval of the MFV-
Tone. Default is ASCII 70
S12
Maximum Supervision Time for an escape
sequence. Default is 0.5 seconds (the time is
given in hundredths of a second)
Note: S6 to S11 are not applicable to ISDN/Digital terminals
T
ABLE
2-15. R
ESULT
C
ODES
N
OT ALL OF THE
H
AYES
R
ESULT
C
ODES ARE USED
. R
ESULT
C
ODES ARE SENT AS REACTIONS TO COMMANDS
OR MESSAGES FROM THE
S
WITCH
. I
F AND WHEN THE
R
ESULT
C
ODES ARE EXPRESSED DEPENDS UPON THE
APPLIED
V
COMMANDS
(V0 & V1)
AND THE
X
COMMANDS
(X0, X1 & X3).
0
OK
Command executed
1
CONNECT
Connected at 0 .. 300 bit/s
2
RING
Incoming call
3
NO CARRIER
No synchronisation or synchronisation lost
4
ERROR
Error in given command
5
CONNECT 1200
Connected at 1200 bit/s
7
BUSY
Busy
10
CONNECT 2400
Connected at 2400 bit/s
11
CONNECT 4800
Connected at 4800 bit/s
12
CONNECT 9600
Connected at 9600 bit/s
13
CONNECT>xxxx
Connected at xxxx bit/s (over 9600 bit/s)
Summary of Contents for ErgoLine D310
Page 1: ...ErgoLine D330 D340 D310 D320 Customer Engineer Manual ...
Page 2: ...3513 302 60092 February 1999 ...
Page 6: ...ErgoLine D310 D320 Customer Engineer Manual ii ...
Page 20: ...ErgoLine D330 D340 Customer Engineer Manual Page 1 14 ...
Page 68: ...ErgoLine D330 D340 Customer Engineer Manual Page 3 4 ...
Page 74: ...ErgoLine D330 D340 Customer Engineer Manual Page 4 6 ...
Page 88: ...ErgoLine D310 320 Customer Engineer Manual Page 5 14 ...