This chapter details how to set up the DreamStation BiPAP autoSV device. This chapter includes:
• Installing the air filter
• Positioning the device
• Connecting the breathing circuit
• Supplying AC power to the device
• Menu and screen navigation
•
Bluetooth
and Wi-Fi setup
2.1 Installing and replacing the air filters
Caution
A properly installed, undamaged Philips Respironics blue pollen filter is required for proper
operation.
The device uses a reusable blue pollen filter that can be rinsed and a disposable light-blue ultra-fine filter.
The reusable blue filter screens out pollens, while the light-blue ultra-fine filter provides more complete
filtration of very fine particles. The reusable blue filter must be in place at all times when the device is
operating. The ultra-fine filter is recommended for people who are sensitive to tobacco smoke or other small
particles.
The reusable blue filter is supplied with the device. A disposable light-blue ultra-fine filter may also be
included. If your filter is not already installed when you receive your device, you must at least install the
reusable filter before using the device.
This device has an automatic air filter reminder. Every 30 days, the device will display a message reminding
you to check your filters and replace them as directed.
Note
This message is a reminder only. The device does not detect the performance of the filters nor does
it recognize when a filter has been rinsed or replaced.
1. Lift up on the filter access door and swing open. If replacing, pull out the old filter assembly.
2. Device setup