Calibration procedure
HoloMonitor should be re-calibrated before use. Image calibration gives a background image
that is used by Hstudio to lower background disturbances and increase the image quality.
The calibration procedure takes approximately one minute.
1
PERIODIC IMAGE CALIBRATION
Start the
calibration procedure by clicking
the
Calibration button
in the
Calibra
-
tion side panel
in the
Live capture tab
in Hstudio.
2
CALIBRATION WIZARD
The three
Diagnostic values
measure
the image quality. All values should be
in the green range. If this is not the case,
please follow the
(page 18). After cleaning, click
Recali
-
brate
to refresh the
Diagnostics values
.
3 CALIBRATION WARNING
If Hstudio fails to perform a calibration, a
warning will appear. After verifying that
HoloMonitor is properly installed and
cleaned, click
OK
.
4 FURTHER PROBLEMS
If any of the
Diagnostic values
remain outside the green range after several cleaning
attempts and overnight drying (see
), contact the technical support at
for further assistance.