Cleaning instructions
1
CLEANING OPTICAL SURFACES
•
The top of the objective (A) and the object beam
window (B) are cleaned with a fresh cotton swab
dipped in 70% ethanol.
•
Use a dry cotton swab to remove ethanol residues.
•
To validate that the optical surfaces are clean,
calibrate HoloMonitor according to the
(page 19), to ensure that the
Diag-
nostic Values
are within the green ranges.
•
Repeat cleaning, if the
Diagnostic Values
are not
in the green ranges.
An instruction video of how to clean the instrument is available at
2
CLEANING INSTRUMENT EXTERIOR
•
The instrument can be wiped down with a damp cloth (ethanol or water). Cleaning must be
performed so that no liquid enters the interior of the instrument.
A
B