ALARMS AND SELF-DIAGNOSIS
This unit has the alarm functions and self-diagnosis functions listed below.
An error code and error message are displayed on the message column when an alarm or self-diagnosis
goes off.
Contact our sales representative or agent if the alarm mode continues, since the failure of the unit is
supposed.
Before calling for service, take some precautions to prevent the stored items (for example by
transferring the items to another refrigerator).
The first letter of the error code means:
W: Warning - The cooling performance is significantly reduced.
A: Alarm - The cooling performance may decline and the temperature of the chamber may rise.
S: Status - There is a possibility of failure other than the cooling performance. Or the status of the unit is
notified.
*1: When the optional battery kit for power failure alarm is installed.
*2: The chamber temperature is displayed for five seconds and the buzzer stops when the buzzer stop
key (BUZZER STOP) is pressed during the power failure alarm. The red LED lamp continues to blink.
Error
code
Error message
Alarm content
LED
indicator
Buzzer
Remote
alarm
Alarm
log
W01
Ref.SNSR Open
Disconnection of thermal
sensor
Blinking of
red lamp
Intermittent
tone
Alarm mode
Logged
W02
Ref.SNSR Short
Short-circuit of thermal
sensor
Blinking of
red lamp
Intermittent
tone
Alarm mode
Logged
W05
Freeze Warning
Alarm of chamber temp.;
equal or lower than 0
o
C
Blinking of
red lamp
Intermittent
tone
Alarm mode
Logged
W06
Ref.Temp.High
High-temperature alarm
Blinking of
red lamp
Intermittent
tone
Alarm mode
Logged
W07
Ref.Temp.Low
Low-temperature alarm
Blinking of
red lamp
Intermittent
tone
Alarm mode
Logged
W10
*1
Power Failure
Power failure alarm
*2
Blinking of
red lamp
Intermittent
tone
Alarm mode
Logged
W12
INT Fan Lock
Chamber fan lock
Blinking of
red lamp
Intermittent
tone
Alarm mode
Logged
W13
Comp. Fan Lock
Compressor fan lock
Blinking of
red lamp
Intermittent
tone
Alarm mode
Logged
A01
Door Open
Door alarm
Blinking of
red lamp
Intermittent
tone
---
---
A02
Def.SNSR Open
Disconnection of defrost
sensor
Blinking of
red lamp
Intermittent
tone
---
Logged
A03
Def.SNSR Short
Short-circuit of defrost
sensor
Blinking of
red lamp
Intermittent
tone
---
Logged
A06
Ref.Temp.High
Delayed status of high-temp.
alarm
Blinking of
red lamp
---
---
---
A07
Ref.Temp.Low
Delayed status of low-temp.
alarm
Blinking of
red lamp
---
---
---
S01
Door Delay ON
Delayed status of door alarm
Blinking of
red lamp
---
---
---
S02
*1
Battery Err.
Battery installation error
Blinking of
red lamp
Intermittent
tone
---
---
S03
*1
Replace Battery
The time for battery
replacement
---
---
---
---
S04
COMM Err.
Communication error
Blinking of
red lamp
---
---
Logged