SETTING THE SELECTABLE PARAMETERS
Setting the remote alarm
The alarm of this unit can be informed at a remote location from this unit by connecting the external alarm
device (commercial goods) to the remote alarm terminals. It is selectable that the operation of the
external alarm device is linked with the buzzer or not.
In the case of link: The alarm of the external device is cancelled when by buzzer stop key (BUZZER
STOP) is pressed during buzzer operation. But the alarm is not cancelled in the case of power failure
alarm.
In the case of no-link: The alarm of the external device is not cancelled when by buzzer stop key
(BUZZER STOP) is pressed during buzzer operation.
Setting range of remote alarm: OFF (no-link), ON (link)
Initial setting (factory setting): OFF (no-link)
1.
On the home screen, press the menu key (MENU/HOME).
►The Menu screen is displayed.
2.
By pressing the up or down key, move the cursor to the Alarms &
Controls [Fig. 1] and then press the enter key (ENTER).
►The Alarms & Controls screen is displayed.
3.
By pressing the up or down key, move the cursor to the Alarm Setting
[Fig. 2] and then press the enter key (ENTER).
►The Alarm Setting screen is displayed.
4.
By pressing the up or down key, move the cursor to the Remote
ON/OFF [Fig. 3] and then press the enter key (ENTER).
►The Remote ON/OFF screen is displayed, and the current setting
(OFF) is displayed [Fig. 4].
5.
By pressing the up or down key, change the setting to ON.
The setting is changed to ON or OFF when the up or down key is
pressed.
6.
Press the enter key (ENTER).
►The setting is memorized and the screen is changed to the Alarm
Setting screen.
7.
Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
[Fig. 1]
[Fig. 2]
[Fig. 3]
[Fig. 4]
Menu
Set Temp.
5
o
C
Keypad Lock
Temp. Max / Min
Data Log
Alarms & Controls
Alarms & Controls
1/10 Temp. Display
Alarm Setting
Light Setting
Comp. Delay
DAQ Setting
Date & Time
Alarm Setting
High Temp.
Low Temp.
Door Alarm Delay
Ring Back Delay
Alarm Volume
Remote ON/OFF
Remote ON/OFF
OFF
ON / OFF