SETTING THE SELECTABLE PARAMETERS
Setting the remote control
By the remote control, the setting of chamber temperature can be changed through the software of the
optional data acquisition system (MTR-5000).
Setting range of remote control:
LOCAL (change of the chamber set temperature remotely is unable) or,
REMOTE (change of the chamber set temperature remotely is enable)
Initial setting (factory setting): LOCAL (change of the chamber set temperature remotely is unable)
1.
On the home screen, press the menu key (MENU/HOME).
►The Menu screen is displayed.
2.
By pressing the up or down key, move the cursor to the Alarms &
Controls [Fig. 1] and then press the enter key (ENTER).
►The Alarms & Controls screen is displayed.
3.
By pressing the up or down key, move the cursor to the DAQ Setting
[Fig. 2] and then press the enter key (ENTER).
►The DAQ Setting screen is displayed.
4.
By pressing the up or down key, move the cursor to the Local /
Remote [Fig. 3] and then press the enter key (ENTER).
►The Local / Remote screen is displayed, and the current setting
(LOCAL) is displayed [Fig. 4].
5.
By pressing the up or down key, change the setting to REMOTE.
The setting is changed to LOCAL or LEMOTE when the up or down
key is pressed.
6.
Press the enter key (ENTER).
►The setting is memorized and the screen is changed to the DAQ
Setting screen.
7.
Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
Note: In the case of REMOTE setting, R is displayed upper right on the
Set Temp. screen. This means the change of the chamber set
temperature is unable.
[Fig. 1]
[Fig. 2]
[Fig. 3]
[Fig. 4]
Menu
Set Temp.
5
o
C
Keypad Lock
Temp. Max / Min
Data Log
Alarms & Controls
Alarms & Controls
1/10 Temp. Display
Alarm Setting
Light Setting
Comp. Delay
DAQ Setting
Date & Time
DAQ Setting
DAQ ID
DAQ Speed
Local / Remote
Local / Remote
LOCAL
LOCAL / REMOTE