Table 7~9 show the behavior of the alarm (buzzer) and Ring Back function when pressing Buzzer key.
Table 7 In the case of other than table 8 or table 9.
Remote Alarm setting
Ring Back
setting
Buzzer from O
2
/CO
2
incubator
Remote Alarm
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
OFF
(Alarm is not
canceled)
ON
ON
ON
(Under continuation)
OFF OFF
OFF: Interlock
with Buzzer key
ON ON
OFF (Alarm is
not canceled)
ON
OFF OFF
OFF
Note:
Resolve the cause of the alarm in reference to page 91~94 because the alarm itself is not
deactivated by pressing Buzzer key.
Table 8 In the case of high limit temperature alarm or Err38 (The outer door opens during H
2
O
2
decontamination; refer to next page).
Remote Alarm setting
Ring Back
setting
Buzzer from O
2
/CO
2
incubator
Remote Alarm
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
ON
ON
(Under continuation)
ON
(Continue)
ON
(Under continuation)
OFF
OFF: Interlock
with Buzzer key
ON
OFF
Note:
Close the outer door when Err38 is activated.
Table 9 In the case of Err01 (CO
2
gas cylinder empty), Err02 (N
2
/O
2
gas cylinder empty), Err11, 12
(CO
2
sensor error), Err18 (UV lamp failure) or door alarm*
2
.
Remote Alarm setting
Ring Back
setting
Buzzer from O
2
/CO
2
incubator
Remote Alarm setting
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
OFF
(Alarm is
canceled)
OFF
(Alarm is already
canceled)
OFF
(Alarm is
canceled*
2
)
OFF
(Alarm is already
canceled*
2
)
OFF
OFF: Interlock
with Buzzer key
ON
OFF
*2: In the door alarm, the remote alarm does not work.
Note:
·
When Err01 is activated, connect the new CO
2
gas cylinder and press the Buzzer key to stop the buzzer.
In addition, when the optional MCO-21GC is installed and the gas supply is switched to the reserve gas
cylinder, press the Buzzer key and replace the gas cylinder.
·
When Err02 is activated, press the Buzzer key and replace the N
2
gas (O
2
gas) cylinder.