86
GAS AUTO CHANGER (OPTION)
Manual CO
2
gas supply line A/B changeover
It is possible to change CO
2
gas supply line A/B manually anytime.
Example) Change CO
2
gas supply line A to B.
1.
Press CO
2
gas supply line select key for a few
seconds.
2.
CO
2
gas supply line A is changed to B.
Note:
The behavior for the following case is shown in Table 5.
After the CO
2
gas supply line A is changed to B by CO
2
gas automatic changer function, the CO
2
gas
supply line B is changed to A manually without pressing the Buzzer key.
Table 5
Situation
CO
2
gas
CO
2
gas supply line indicator
Message
display field
Supply line Cylinder A Cylinder B
A
B
1
CO
2
gas supply line A
is changed to B
automatically.
B Empty
Remaining
Reverse
video
and blink
Light on
Err01:
CO
2
Gas Empty
(and buzzer)
2
Without pressing the
Buzzer key,
long-pressed CO
2
gas supply line select
key.
A Empty
Remaining
Blink Light
off
Err01:
CO
2
Gas Empty
(and buzzer)