On the Alarm Setting #1 screen, input each
parameter. Press Apply key to save the input value.
The display returns to the Tools #1 screen.
●
Each parameter setting
·
Automatic set temperature alarm (Temp. Alarm):
When the chamber temperature exceeds the scope, the set temperature ± the set value of Automatic set
temperature alarm, the alarm is activated. Settable range: ±1.0
o
C~±5.0
o
C, factory setting: ±1.0
o
C.
·
Automatic set temperature alarm delay (Temp. Alarm Delay):
The function is that when the incubator is in the alarm state of automatic set temperature, the alarm buzzer
will sound after the automatic set temperature alarm delay set time passed.
Settable range: 0 minute~15 minutes, factory setting: 15 minutes.
Note:
When the incubator is recovered from the alarm state of automatic set temperature within the
automatic set temperature alarm delay time, the buzzer doesn’t sound after the elapse of the automatic set
temperature alarm delay.
·
Automatic set CO
2
density alarm (CO
2
Alarm):
When the chamber CO
2
density exceeds the scope, the set CO
2
density ± the set value of automatic set
CO
2
density alarm, the alarm is activated. Settable range: ±0.5 %~±5.0 %, factory setting: ±1.0 %.
·
Automatic set CO
2
density alarm delay (CO
2
Alarm Delay):
The function is that when the incubator is in the alarm state of automatic set CO
2
density, the alarm buzzer
will sound after the automatic set CO
2
density alarm delay set time passed.
Settable range: 0 minute~15 minutes, factory setting: 15 minutes.
Note:
When the incubator is recovered from the alarm state of automatic set CO
2
density within the
automatic set CO
2
density alarm delay time, the buzzer doesn’t sound after the elapse of the automatic set
CO
2
density alarm delay.
·
Automatic set O
2
density alarm (O
2
Alarm):
When the chamber O
2
density exceeds the scope, the set O
2
density ± the set value of automatic set O
2
density alarm, the alarm is activated. Settable range: ±0.5 %~±5.0 %, factory setting: ±1.0 %.
·
Automatic set O
2
density alarm delay (O
2
Alarm Delay):
The function is that when the incubator is in the alarm state of automatic set O
2
density, the alarm buzzer
will sound after the automatic set O
2
density alarm delay set time passed.
Settable range: 0 minute~60 minutes, factory setting: 30 minutes.
Note:
When the incubator is recovered from the alarm state of automatic set O
2
density within the
automatic set O
2
density alarm delay time, the buzzer doesn’t sound after the elapse of the automatic set
O
2
density alarm delay.