CALIBRATION:
The 9500 series indicators have a built‐in calibration feature which should only be used
for fine adjustments. If the obtained results on your calibration blocks are well outside of
tolerance, you should adjust and repair the base machine before you adjust the indicator.
1) Press the Menu/Enter button. Scroll down to System and highlight it. Press the
Menu/Enter button again and you must scroll down to Correction.
2) Press the menu/Enter button and you will see three choices:
a) Off
b) On
c) Enter
If you don’t want to change calibration then leave it in the off position and return to the
main test screen.
If you highlight ON, this will activate the Auto‐Calibration feature which will allow the
indicator to automatically correct the measured value when the “Lock” feature is on.
If you highlight Enter, this will allow you to set the calibration by manually entering in the
value shown on the test block.
Example: If the value shown on your test block is HRC 25.8 then you would enter that
value while in the enter screen.
Now take a test on that same test block as normal.
Once the result has been obtained Press the Menu/Enter button and then choose Save and
then save the calibration. If you choose Exit then the calibration will not be saved.
Maintenance:
The 9500 series indicators are highly sensitive and accurate instruments and should be
treated as such.
These should remain in a temperature controlled environment with the least amount of
shock or vibration as possible.
Cleaning:
Use soft damp cloth only. Do not use harsh chemicals as this will deteriorate the finish of
the indicator.