12
Installation
Fig. 6: Example for fuse protection for single phase operation
max. current: 17.7 A
Changing the voltage
range
The mains voltage must be determined on-site each time before the pump is in-
stalled or moved to a different location.
A voltage selector switch below the connector plate enables toggling between the
two voltage ranges. The factory default setting is 200 ... 240 VAC.
Fig. 7: Changing the voltage range for the power supply
Disconnect the pump from the power supply.
L1
N
PE
02
04
01
F 1
NOTE
Overvoltage!
An incorrect voltage range setting can damage the motor.
Disconnect the pump from the power supply.
Only change the voltage range when the pump is disconnected from the power
mains.
S
Voltage selector switch
240
240
120
S