Display Problem
Possible causes
Remedy
E698
Communication
error
● Electronic drive unit not responding
● Contact Pfeiffer Vacuum Service.
F110
Pressure gauge
● Transmitter faulty
● Connection to the transmitter discon-
nected during operation
● Restart with connected measuring tube
● Replace the pressure measuring tube
● Install the pressure measuring tube cor-
rectly
Tbl. 11:
Error displays and warnings for the DCU display and control unit
Display Problem
Possible causes
Remedy
E001
Overspeed
● Call Pfeiffer Vacuum Service
● Only acknowledge for rotational speed f = 0
E002
Overvoltage
● Incorrect power pack
used
● Check power pack type
● Check partial mains voltage
E006
Run-up fault
● Run-up time threshold
set too low
● Gas flow in recipient
through leaks or open
valves
● Speed-control switching
point not reached upon
expiration of run-up time
● Adjust run-up time to process conditions
● Check recipient for leakage and closed
valves
● Adjust speed-control switching point
E007
Insufficient operating fluid
● Insufficient operating
fluid
● Check operating fluid
● Only acknowledge for rotational speed f = 0
E008
Connection from electronic
drive unit to pump faulty
● Connection to pump
faulty
● Check connections
● Only acknowledge for rotational speed f = 0
E010
Internal device error
● Call Pfeiffer Vacuum Service
● Only acknowledge for rotational speed f = 0
E021
Drive electronics fail to identi-
fy pump
● Call Pfeiffer Vacuum Service
● Only acknowledge for rotational speed f = 0
E043
Internal configuration error
● Call Pfeiffer Vacuum Service
E044
Excess temperature electron-
ics
● Insufficient cooling
● Improve cooling
● Check deployment conditions
E045
Motor overheated
● Insufficient cooling
● Improve cooling
● Check deployment conditions
E046
Internal initialization error
● Call Pfeiffer Vacuum Service
E091
Internal device error
● Call Pfeiffer Vacuum Service
E092
Unknown terminal panel
● Call Pfeiffer Vacuum Service
E093
Temperature evaluation on
motor is faulty
● Call Pfeiffer Vacuum Service
E094
Temperature evaluation on
electronics is faulty
● Call Pfeiffer Vacuum Service
E098
Internal communication error
● Call Pfeiffer Vacuum Service
E107
Combined error for output
stage
● Call Pfeiffer Vacuum Service
● Only acknowledge for rotational speed f = 0
E108
Speed measurement fault
● Call Pfeiffer Vacuum Service
● Only acknowledge for rotational speed f = 0
E109
Software not released
● Call Pfeiffer Vacuum Service
E110
Operating material analysis
contains errors
● Call Pfeiffer Vacuum Service
● Only acknowledge for rotational speed f = 0
Malfunctions
41/56