Risks in the event of faults
WARNING
Danger of death due to the turbopump tearing off in the event of a fault
Sudden blocking of the rotor generates high destructive torques in accordance with ISO 27892. If it is
not
fastened correctly, the turbopump will tear off. The energy released in this way can propel the
entire pump or broken pieces from the inside of the pump through the room. Potentially dangerous
gases may escape. There is a risk of serious injury, potentially even fatal, and significant equipment
damage.
► Follow the installation instructions from the manufacturer.
► Observe the requirements for stability and the design of the counter flange.
► Only use approved original accessories or mounting elements approved by Pfeiffer Vacuum for
the installation process.
WARNING
Danger of death due to the turbopump with vibration damper is tearing off in the event of a
fault
Sudden blocking of the rotor generates high destructive torques in accordance with ISO 27892.
These result in the turbopump tearing off when using a vibration damper. The energy released in this
way can throw the entire pump or broken pieces from the inside of the pump through the space. Po-
tentially dangerous gases may escape. There is a risk of serious injury, potentially even fatal, and
significant property damage.
► Provide suitable safety measures on customer's side to compensate the torques arising.
► Please contact Pfeiffer Vacuum before installing a vibration damper.
2.3 Safety precautions
Information requirement on possible hazards
The owner or operator of the product is required to make all operating personnel
aware of hazards arising from the use of this product.
Every person who is involved in the installation, operation or maintenance of the
product must have read and understood the safety-relevant parts of this documen-
tation.
Infringement of EU conformity due to modifications to the product
The Declaration of Conformity from the manufacturer is no longer valid if the opera-
tor changes the original product or installs additional equipment.
● Following installation into a system, the operator is required to check and re-
evaluate as necessary the conformity of the overall system in the context of the
relevant EU Directives before commissioning that system.
General safety measures when handling the product
► Observe all applicable safety and accident prevention regulations.
► Check that all safety measures are observed at regular intervals.
► Do not expose any body parts to the vacuum.
► Always ensure a safe connection to the protective conductor (PE).
► Do not unplug any plug connectors during operation.
► Observe the specified shutdown procedures.
► Wait for the rotor to come to a complete standstill (speed f = 0) before working on the high vac-
uum connection.
► Do not put the device into operation with an open high vacuum flange.
► Keep lines and cables away from hot surfaces (> 70 °C).
► Never fill or operate the device with cleaning agents or cleaning agent residue.
► Do not carry out any unauthorized modifications or changes to the device.
► Observe the protection rating of the device being installation or operation in different environ-
ments.
Safety
11/56